1
00:00:45,640 --> 00:00:50,890
And dissolving it into tetrahydrofuran
so this is a complete dissolution the

2
00:00:50,890 --> 00:00:58,600
plastic will dissolve and we now have a
solution if you Sonic eight this sample

3
00:00:58,600 --> 00:01:05,470
while the when you first put the PVC toy
in typically with the 1530 minute

4
00:01:05,470 --> 00:01:07,540
sonication
you'll have a completely dissolve

5
00:01:07,540 --> 00:01:11,380
solution so it makes things a little bit
quicker after we have a dissolve

6
00:01:11,380 --> 00:01:16,390
solution will precipitate in hexanes
this removes a PVC from the solution so

7
00:01:16,390 --> 00:01:20,920
we'll get a solid precipitate that
leaves a phalates in solution makes it a

8
00:01:20,920 --> 00:01:25,290
little bit easier to analyze so we'll
move into the analysis step with GCMs

9
00:01:25,290 --> 00:01:30,790
and I'll mention here that the cpsc
method allows for several alternative

10
00:01:30,790 --> 00:01:35,979
methods within it this is too important
for testing labs of labs already

11
00:01:35,979 --> 00:01:40,630
familiar or likes to use a certain test
method that method those that are listed

12
00:01:40,630 --> 00:01:45,720
are available to use instead so
generally the alternative methods

13
00:01:45,720 --> 00:01:50,500
usually what differs is the extraction
technique and most of them fall into one

14
00:01:50,500 --> 00:01:54,670
of the categories I have listed here a
couple of them fall into just a typical

15
00:01:54,670 --> 00:01:59,500
fluid extraction where you're cutting up
the sample and you're putting it into a

16
00:01:59,500 --> 00:02:05,619
solvent usually dichloromethane and
letting it sit the California Department

17
00:02:05,619 --> 00:02:12,880
of toxic substances method asked just
for a 15 minute time and the Health

18
00:02:12,880 --> 00:02:18,340
Canada method is much longer rate
12-hour time so even with this with this

19
00:02:18,340 --> 00:02:24,090
step you have a long range of sample
times Sox flu extractions a more of a

20
00:02:24,090 --> 00:02:28,690
aggressive fluid extraction here you
have an apparatus that I have shown on

21
00:02:28,690 --> 00:02:34,750
the right where your you have a
ground-up sample and the solvent is is

22
00:02:34,750 --> 00:02:38,680
heated and then cool the solvent will
come down through the sample so you're

23
00:02:38,680 --> 00:02:43,570
having a continuous extraction with
fresh solvent usually the the methods

24
00:02:43,570 --> 00:02:48,280
call for about six hours methods under
this category are the the Chinese method

25
00:02:48,280 --> 00:02:52,810
and European methods which I have listed
and also EPA methods there's some other

26
00:02:52,810 --> 00:02:57,250
older methods that we we see using this
as well there are a couple other

27
00:02:57,250 --> 00:03:00,690
techniques
typically found with within the EPA

28
00:03:00,690 --> 00:03:07,080
methodologies they were looking at using
different tools for your extraction such

29
00:03:07,080 --> 00:03:13,860
as using a microwave pressurized fluid
extraction or ultrasonic extraction and

30
00:03:13,860 --> 00:03:20,010
the common bond here we're looking for
is an analysis by GCMs we think this is

31
00:03:20,010 --> 00:03:24,300
important g CMS's are our preferred
analysis technique because it's

32
00:03:24,300 --> 00:03:31,140
qualitative and quantitative here I have
a chromatogram shown of an analysis with

33
00:03:31,140 --> 00:03:37,670
a standard of all six of the
specifically regulated phalates and

34
00:03:37,670 --> 00:03:43,019
generally we're getting a pretty good
separation especially for the lower

35
00:03:43,019 --> 00:03:50,670
molecular weight ones and obviously the
dno PDI NP + di DP which you see in the

36
00:03:50,670 --> 00:03:56,010
later half the chromatogram alone around
11 12 minutes that's where things get a

37
00:03:56,010 --> 00:04:00,000
little bit trickier and we'll be talking
a little bit more about that today we

38
00:04:00,000 --> 00:04:06,780
specified GCMs because some older
methods call for GCF ID and simply put

39
00:04:06,780 --> 00:04:09,810
if you're having any overlap and you're
chromatogram with different with

40
00:04:09,810 --> 00:04:14,670
different phalates f ID is just not
strong enough to get a full qualitative

41
00:04:14,670 --> 00:04:21,840
and quantitative assessment so I'm gonna
briefly go over some common problems I'm

42
00:04:21,840 --> 00:04:25,770
gonna give you very very simple problems
and very simple solutions that probably

43
00:04:25,770 --> 00:04:31,260
don't do these justice but just to get a
the the ideas flowing for discussion

44
00:04:31,260 --> 00:04:38,100
later some common questions we get often
people are asked is there a standard

45
00:04:38,100 --> 00:04:45,810
reference material for phalates in PVC
another common problem is false positive

46
00:04:45,810 --> 00:04:51,990
results many of manufacturers someone
else might say I did not put phalates in

47
00:04:51,990 --> 00:04:56,160
this item I'm sure there's no phalates
in this item but the test lab still said

48
00:04:56,160 --> 00:05:01,979
that that there are phalates in this so
that could either come under a false

49
00:05:01,979 --> 00:05:07,370
positive result or a result of
contamination somewhere along the lines

50
00:05:07,370 --> 00:05:11,910
another issue is the identity of dice
and Nona land dice adecco Fowley

51
00:05:11,910 --> 00:05:17,340
both of these phalates actually have
multiple cast numbers so there's

52
00:05:17,340 --> 00:05:22,080
different ways of making these compounds
and there's a different chemical formula

53
00:05:22,080 --> 00:05:29,100
of these compounds so try it to trying
to summarize some of these issues and

54
00:05:29,100 --> 00:05:36,330
solutions as best we can
there are multiple issues of the cpsc

55
00:05:36,330 --> 00:05:40,710
method it's been updated a few times
we're now on the third version sometimes

56
00:05:40,710 --> 00:05:45,720
people with problems they'll contact us
and say well I was following the first

57
00:05:45,720 --> 00:05:50,190
version of this and this is a problem
I've had there are there are significant

58
00:05:50,190 --> 00:05:54,150
advancements between the versions so
make sure you're definitely using the

59
00:05:54,150 --> 00:06:01,020
most current one we have an interagency
agreement with NIST to develop standard

60
00:06:01,020 --> 00:06:06,690
reference materials this has been an
ongoing project the last update has been

61
00:06:06,690 --> 00:06:11,370
that they're hoping to have something
within the next calendar year we'll see

62
00:06:11,370 --> 00:06:17,970
how that develops but this is a
promising idea there's also some

63
00:06:17,970 --> 00:06:21,680
companies that we've seen I believe
specs Serta prep has started issuing a

64
00:06:21,680 --> 00:06:27,890
certified reference material they have a
specific level of phalates in

65
00:06:27,890 --> 00:06:32,580
polyethylene so that's another frontier
you can look with standards companies

66
00:06:32,580 --> 00:06:36,720
they may also be developing their own or
looking for something along those lines

67
00:06:36,720 --> 00:06:43,170
I mentioned contamination before this is
possible and this is a good question to

68
00:06:43,170 --> 00:06:48,360
ask if you're a manufacturer or seller
essentially from the starting points at

69
00:06:48,360 --> 00:06:51,960
the endpoint have I looked at every
possible area where phalates can come

70
00:06:51,960 --> 00:06:57,690
into my line we've heard of recycling
plastics sometimes people use recycled

71
00:06:57,690 --> 00:07:02,040
plastics that had fights and before and
that could that level of phalates and

72
00:07:02,040 --> 00:07:09,270
the new material could be enough to lead
to a failure also is the equipment

73
00:07:09,270 --> 00:07:12,210
that's being used and they've run
phalates and they've been properly

74
00:07:12,210 --> 00:07:18,090
cleaned essentially you need to break
down and look at every step are you sure

75
00:07:18,090 --> 00:07:23,610
that every step of the way is there's no
way that contamination can come in to

76
00:07:23,610 --> 00:07:29,180
this issue
and finally a lot of our suggestions and

77
00:07:29,180 --> 00:07:33,980
emphasis is on qualitative analysis by
technical staff and I'll be going into

78
00:07:33,980 --> 00:07:41,300
this more but we have a few key points
here we are CPSC method calls for a

79
00:07:41,300 --> 00:07:45,050
retention time match with a standard
phalates so make sure when you're

80
00:07:45,050 --> 00:07:48,470
looking at the chromatogram and the
chromatogram of a sample a retention

81
00:07:48,470 --> 00:07:53,510
time must match or else that should be a
red flag also a mass spectrum match this

82
00:07:53,510 --> 00:07:57,710
and not just a general mass spectrum but
the iron ratio is very important and can

83
00:07:57,710 --> 00:08:03,710
help in identifying blank analysis is
crucial for looking for contamination

84
00:08:03,710 --> 00:08:08,030
within the lab result itself if you run
a blank and you're seeing phalates in

85
00:08:08,030 --> 00:08:13,100
there that means you have contamination
along somewhere along the line if people

86
00:08:13,100 --> 00:08:18,710
are reusing glassware we found in our
lab that our GCMs Inlet liner sometimes

87
00:08:18,710 --> 00:08:23,960
can carry these fat lights over so it's
they're persistent enough that they can

88
00:08:23,960 --> 00:08:30,020
come along the line even within the lab
and I mentioned before the ein P and D

89
00:08:30,020 --> 00:08:35,360
IDP have multiple caste numbers if you
look closely new and you get standards

90
00:08:35,360 --> 00:08:39,590
of both you'll see there chromatogram
peak shape is a little different so this

91
00:08:39,590 --> 00:08:43,850
is a great way to identify which
material you're working for and you

92
00:08:43,850 --> 00:08:49,520
should be if you're quantifying one
specific caste number of di MP or di DP

93
00:08:49,520 --> 00:08:55,580
your standard should be that same caste
number so you get a fully well developed

94
00:08:55,580 --> 00:09:01,580
match so I'll just give you a few
examples of why a strong qualitative

95
00:09:01,580 --> 00:09:07,040
assessment I'll help here I have two
chromatograms one is of di NP and the

96
00:09:07,040 --> 00:09:12,260
other is of a common Fallot alternative
din CH if you're looking at the peak

97
00:09:12,260 --> 00:09:15,680
shape they mostly look the same if
you're looking specifically at peaks and

98
00:09:15,680 --> 00:09:20,390
valleys you can kind of tell they're a
little different but our experience and

99
00:09:20,390 --> 00:09:25,040
I can see this translating to other labs
is if you have a sort of an automated

100
00:09:25,040 --> 00:09:28,280
quantification procedure where the
computer will sort of do the work for

101
00:09:28,280 --> 00:09:33,140
you if it's looking for a broad peak in
this time range it may not know the

102
00:09:33,140 --> 00:09:39,050
difference between the InP and di di n
CH so here this is where

103
00:09:39,050 --> 00:09:45,080
mass spectrum look can really illusi
eight the differences the Dench base ion

104
00:09:45,080 --> 00:09:52,550
is 155 vs di NP which is 149 so all
really takes is a quick look at the mass

105
00:09:52,550 --> 00:09:56,290
spectrum and you'll be able to tell the
difference

106
00:09:56,330 --> 00:10:01,520
another issue we've run into is DN o P
vs. dioctyl terephthalate another common

107
00:10:01,520 --> 00:10:07,130
alternative here I have chromatograms of
the two of those you can see the

108
00:10:07,130 --> 00:10:11,890
retention time is awfully close 11.15
minutes versus eleven point 100 minutes

109
00:10:11,890 --> 00:10:18,730
and here I've noticed that our our
previous settings for our computer

110
00:10:18,730 --> 00:10:24,620
quantification program if the tolerances
are empty enough or are wide enough

111
00:10:24,620 --> 00:10:28,130
sorry it may not be able to tell the
difference especially since their math

112
00:10:28,130 --> 00:10:31,520
spectrum is pretty similar if you're
looking if you're running a sim scan and

113
00:10:31,520 --> 00:10:41,570
looking at the typical DN o P Mass ions
do T P covers those as well but if you

114
00:10:41,570 --> 00:10:45,920
do if you're thorough and follow through
with your qualitative assessment you'll

115
00:10:45,920 --> 00:10:51,740
see here's our DN o P a typical mass
spectrum they speak again is 149 there's

116
00:10:51,740 --> 00:10:58,160
a little bit of 167 and 279 mass ions
we'll compare that directly with do T P

117
00:10:58,160 --> 00:11:04,520
and here you can see the the jump and
the 261 ion it's almost 50% relative to

118
00:11:04,520 --> 00:11:10,880
149 so if you're looking if you're again
if you're doing an automated program you

119
00:11:10,880 --> 00:11:20,570
might get a false positive here if you
had do T P so okay so we'll start to

120
00:11:20,570 --> 00:11:25,280
shift into the the rest of the talk here
or the rest of the symposium with the

121
00:11:25,280 --> 00:11:29,870
discussion about emerging technologies
we'll have a block of speakers today

122
00:11:29,870 --> 00:11:34,700
discussing portable spectroscopy we've
started using this as a valley screening

123
00:11:34,700 --> 00:11:43,730
tool a general kind of yes or no might
this item have phalates it's when this

124
00:11:43,730 --> 00:11:50,120
law was passed people commonly asked
well what's the XRF i'll so if we're

125
00:11:50,120 --> 00:11:56,089
looking for an X or F for lead what's
the fal 8 comparable option we have and

126
00:11:56,089 --> 00:12:02,959
this is so far the most promising
technology also I'll refer you to this a

127
00:12:02,959 --> 00:12:08,329
paper by Zhang at all last year from
chemical communications they developed a

128
00:12:08,329 --> 00:12:12,529
colorimetric test it's a different
application they're looking for phalates

129
00:12:12,529 --> 00:12:17,899
in water but I imagine with the right
ingenuity and the right research this

130
00:12:17,899 --> 00:12:22,880
might be adapted for toys or childcare
articles they use modified gold

131
00:12:22,880 --> 00:12:27,560
nanoparticles that actually changed
color and water when my lights were

132
00:12:27,560 --> 00:12:32,899
present and also we're big fans of
direct analysis in real time mass

133
00:12:32,899 --> 00:12:37,399
spectrometry I will have a talk on that
later so I won't steal too much of that

134
00:12:37,399 --> 00:12:42,290
thunder but it's promising for a
screening tool and we think with the

135
00:12:42,290 --> 00:12:47,360
right research possibly also
qualification method so just a couple

136
00:12:47,360 --> 00:12:52,610
more sources for information we have
CPSC has its own phalates website where

137
00:12:52,610 --> 00:12:57,620
you can get information common questions
also we have a lab frequently asked

138
00:12:57,620 --> 00:13:01,009
questions page there if you're
interested more in the chronic hazard

139
00:13:01,009 --> 00:13:05,870
advisory panel there's also a website
for that they should be issuing a ruling

140
00:13:05,870 --> 00:13:11,930
I believe in that within the calendar
year there's currently an ASTM workgroup

141
00:13:11,930 --> 00:13:17,870
focused simply on phalion analysis for
low level percentages so I have this

142
00:13:17,870 --> 00:13:22,939
information here and finally EPA has a
design for an Environment Program and

143
00:13:22,939 --> 00:13:25,000
they're looking at alternatives to
phalates

144
00:13:25,000 --> 00:13:29,839
generally the idea is anyone that's
interested in the valley realm and

145
00:13:29,839 --> 00:13:34,189
they're interested in alternatives they
can propose alternatives I think they'll

146
00:13:34,189 --> 00:13:38,720
be looking at some toxicology issues so
any company that's looking to move this

147
00:13:38,720 --> 00:13:47,900
might be a good website to follow and
with that I'll take any questions yeah

148
00:13:47,900 --> 00:13:53,420
oh so he asks what NIST is this is the
National Institute for Standards and

149
00:13:53,420 --> 00:14:06,020
Technology yes we consider screening
just essentially a tool or a method to

150
00:14:06,020 --> 00:14:11,300
kind of give you a generic yes-or-no
answer if phalates are present when we

151
00:14:11,300 --> 00:14:14,750
look as a screening tool for if we're
looking for lead we use x-ray

152
00:14:14,750 --> 00:14:19,490
fluorescence as a screening tool this
will give us like essentially 30 seconds

153
00:14:19,490 --> 00:14:24,050
to a minute we have a general idea of if
light is present or there so that's sort

154
00:14:24,050 --> 00:14:38,779
of the itinerary we're looking for with
valleys he asked if there's a acceptable

155
00:14:38,779 --> 00:14:43,550
level for screening as far as false
positives or not I think that's to be

156
00:14:43,550 --> 00:14:47,960
determined you know we're trying to work
with its each technology will have its

157
00:14:47,960 --> 00:14:52,520
own limitations and you know I think the
best we can get on any regards will

158
00:14:52,520 --> 00:14:57,650
we'll work with for now I mean for a
generic tool if you know are the limits

159
00:14:57,650 --> 00:15:02,600
0.1% and I think for a final screening
tool it will be able to it'll need to be

160
00:15:02,600 --> 00:15:07,850
able to reach that but there might be
different ways to adapt back technology

161
00:15:07,850 --> 00:15:16,400
or different applications to try and use
that to your best advantage hi I'm Joel

162
00:15:16,400 --> 00:15:19,850
rectum the chemistry division director
here I just wanted to add to that the

163
00:15:19,850 --> 00:15:24,529
the what's needed from a screening tool
depends somewhat on what it's being used

164
00:15:24,529 --> 00:15:34,029
for as well so affirm we have a speaker
here from BASF today BASF produces

165
00:15:34,029 --> 00:15:41,660
phthalates there need to screen for four
products you know for perhaps impurities

166
00:15:41,660 --> 00:15:45,380
in their line are different than the
needs that we have as a regulatory

167
00:15:45,380 --> 00:15:50,750
agency or a manufacturer who might look
to do screening in between their full

168
00:15:50,750 --> 00:15:56,120
tests so you know there can still be
room for for these screening methods

169
00:15:56,120 --> 00:16:02,440
where it makes sense in a particular end
use

170
00:16:08,380 --> 00:16:29,740
I'm sorry oh yeah we we consider screen
printing applicable to regulation yes

171
00:16:37,410 --> 00:16:43,829
well I think we're we'd like to work
towards that point it would be ideal but

172
00:16:43,829 --> 00:16:47,899
we don't think it's at that point yet

173
00:16:55,429 --> 00:17:02,039
yeah we think they've been used for oh
I'm sorry has do TPD a little

174
00:17:02,039 --> 00:17:08,039
terephthalate been in use often yes we
we have we've we're familiar with it we

175
00:17:08,039 --> 00:17:11,509
think it's been used and used for a
while

176
00:17:15,449 --> 00:17:19,679
that's not my I'm not I can't answer
that that that would be a good question

177
00:17:19,679 --> 00:17:26,629
for the EPA design for alternatives and
and the chronic hazard advisory panel

178
00:17:28,100 --> 00:17:33,870
all right great why don't we move along
our next speaker is Raphael Costa from

179
00:17:33,870 --> 00:17:36,320
Agilent

180
00:17:45,590 --> 00:17:51,270
thank you
thank you Matt for inviting me here to

181
00:17:51,270 --> 00:17:55,020
this symposium is very exciting for us
to be here and what I'm gonna do today

182
00:17:55,020 --> 00:17:59,100
is share with you some of the different
mass spectrometry techniques that may

183
00:17:59,100 --> 00:18:06,050
aid you in solving that problem of
contamination or library matching or

184
00:18:06,050 --> 00:18:10,770
retention times matching up what are
some of the tools that we could use and

185
00:18:10,770 --> 00:18:14,040
what are some of the applications that
admins are our scientists in the field

186
00:18:14,040 --> 00:18:19,170
have developed to meet this growing need
of phthalates and consumer products and

187
00:18:19,170 --> 00:18:24,750
specific in toys and in particular so
I'll go through some of the GCMs

188
00:18:24,750 --> 00:18:29,340
portfolio the different technologies of
different flavors of mass spectrometry

189
00:18:29,340 --> 00:18:34,170
that could aid you in identifying
phthalates in a way that gives you that

190
00:18:34,170 --> 00:18:37,770
much more confidence in your lab I'll go
throughout the application notes

191
00:18:37,770 --> 00:18:41,190
developed by our scientist in the field
and then we'll go through some of the

192
00:18:41,190 --> 00:18:45,000
ongoing work that we internally at
Agilent are working with to solve this

193
00:18:45,000 --> 00:18:49,740
problem of trying to reach the levels
required and the different technologies

194
00:18:49,740 --> 00:18:54,930
and level of confidence in your results
so for example the Agilent GCMs our

195
00:18:54,930 --> 00:18:59,970
portfolio includes not only single
quartz ion traps mobile mass

196
00:18:59,970 --> 00:19:06,200
spectrometers as well as high-end triple
quads and now cute off technology for

197
00:19:06,200 --> 00:19:10,500
mass spectrometry what I will do is I'll
say how each of these different

198
00:19:10,500 --> 00:19:15,210
techniques apply to your particular
analysis what are some of the extra

199
00:19:15,210 --> 00:19:19,830
information that you get from each
individual technique so for example we

200
00:19:19,830 --> 00:19:25,560
know we have here the usual suspects of
the phthalates of interest but their

201
00:19:25,560 --> 00:19:29,600
caste number their molecular weights and
there are corresponding ions for sim

202
00:19:29,600 --> 00:19:35,400
analysis the typical GC conditions that
was used in this particular application

203
00:19:35,400 --> 00:19:42,660
was a very mirror column 0.25 0.25
micron db5 ultra inner column that this

204
00:19:42,660 --> 00:19:46,590
will allow allow you to have more the
ultra nurnies allows you to have more

205
00:19:46,590 --> 00:19:51,480
runs per sample
let's bleed in your chromatograms since

206
00:19:51,480 --> 00:19:57,660
our mass specs are very sensitive there
is a split ratio of 20 to 1 to avoid

207
00:19:57,660 --> 00:20:01,890
contamination
and column overload carrier gases helium

208
00:20:01,890 --> 00:20:06,840
however we're noticing that customers
are asking us hey can we use hydrogen

209
00:20:06,840 --> 00:20:11,340
for analysis and that's something that
you definitely can't do with your single

210
00:20:11,340 --> 00:20:17,280
quads GCMs however we're trying to see
how that could aid you more but the cost

211
00:20:17,280 --> 00:20:21,030
of helium as you may or may not know is
going up drastically recently and we're

212
00:20:21,030 --> 00:20:24,450
trying to address that in terms of uh
how can we take advantage of the

213
00:20:24,450 --> 00:20:28,830
properties of hydrogen as carrier gas
which is not a new thing in Europe they

214
00:20:28,830 --> 00:20:31,920
currently use
that's the carrier gas of choice hope

215
00:20:31,920 --> 00:20:37,050
and we're trying to make our us
customers aware of this as well not only

216
00:20:37,050 --> 00:20:42,150
do you get the benefits of maybe having
a reducing your cost of analysis by

217
00:20:42,150 --> 00:20:46,320
having hydrogen through a hydrogen
generator in your lab but also have

218
00:20:46,320 --> 00:20:50,040
faster chromatography so we definitely
want to ensure that you're aware of

219
00:20:50,040 --> 00:20:54,330
though of that the possibility that that
could bring to your labs well for

220
00:20:54,330 --> 00:21:00,690
example here we have a total iron
chromatogram of a pacifier extract on

221
00:21:00,690 --> 00:21:05,310
spiked and spiked with two parts per
million phthalic mixture on top you see

222
00:21:05,310 --> 00:21:10,830
the mutilated hydroxy toluene as the
main peak in the pacifier however when

223
00:21:10,830 --> 00:21:16,800
we spiked it at 2 ppm we're still able
to see this in matrix uh very nicely so

224
00:21:16,800 --> 00:21:23,430
again not the DI MP and the DI p di DP
are very uh a little bit less sensitive

225
00:21:23,430 --> 00:21:29,970
to particular analysis of VII however
there are things that that are mentioned

226
00:21:29,970 --> 00:21:34,740
throughout my presentation that may
allow you to see higher levels of these

227
00:21:34,740 --> 00:21:40,920
particular compás and we'll go through
that as well okay so for example you're

228
00:21:40,920 --> 00:21:44,190
talking about sample cleanup and
contamination one possible way to avoid

229
00:21:44,190 --> 00:21:49,440
contamination may be to use the SPME or
solid phase microextraction in your

230
00:21:49,440 --> 00:21:53,820
analysis which reduces some of the the
non analytes of interest of all the

231
00:21:53,820 --> 00:21:57,930
matrix of your particular toys or
analysis that would allow you to have

232
00:21:57,930 --> 00:22:02,130
cleaner chromatograms and less
contamination and more throughput per

233
00:22:02,130 --> 00:22:08,160
samples went through GCMs okay so for
example so these for these particular

234
00:22:08,160 --> 00:22:11,670
compounds of interest these phthalates
of interest we are able to regional of

235
00:22:11,670 --> 00:22:15,960
in the level
of from 0.25 milligrams per litre to 10

236
00:22:15,960 --> 00:22:20,730
with a linear dynamic range and we have
individual retention times for these so

237
00:22:20,730 --> 00:22:26,029
we could definitely identify them not
only based on retention time but also on

238
00:22:26,029 --> 00:22:30,960
their spectrum as well with the GCMs one
of the things that you may be that you

239
00:22:30,960 --> 00:22:34,110
may have is if you're doing high
throughput analysis in your lab and you

240
00:22:34,110 --> 00:22:37,940
want to ensure have one method through
all your lab you want to ensure that you

241
00:22:37,940 --> 00:22:43,230
implore something that we call retention
time locking how many of you are used

242
00:22:43,230 --> 00:22:48,059
GCMs for analysis of valets today in
your labs how many of you have an

243
00:22:48,059 --> 00:22:52,440
Agilent system in your lab and how many
of you are using retention time locking

244
00:22:52,440 --> 00:22:56,519
and your instruments so that each
retention time in each instrument that

245
00:22:56,519 --> 00:23:02,369
you have with the same method is exactly
the same one okay I'll talk through that

246
00:23:02,369 --> 00:23:05,999
I'll go through that in my presentation
and how that could aid you and having

247
00:23:05,999 --> 00:23:10,080
that much more confidence it's one of
the hidden secrets that we have and we

248
00:23:10,080 --> 00:23:13,139
love my job here today is to make you
aware of these features that may make

249
00:23:13,139 --> 00:23:16,529
your life a little easier in the lab
when you're analyzing phthalates or

250
00:23:16,529 --> 00:23:24,600
other compounds of interest as well okay
so I'll go here up so for example let's

251
00:23:24,600 --> 00:23:27,960
say you want to take yours your lab
method and you want to take it in the

252
00:23:27,960 --> 00:23:32,009
field in the back of a mobile lab and
you want to test at the point of sale of

253
00:23:32,009 --> 00:23:36,389
our particular product you could do that
with our agilent of 50 975 t for

254
00:23:36,389 --> 00:23:40,679
transportable which implores LTM
technology essentially what you have now

255
00:23:40,679 --> 00:23:44,519
here is you have the same mass
spectrometer that you're used to in your

256
00:23:44,519 --> 00:23:48,450
lab and you're able to take it in the
field at 1/3 of the footprint and half

257
00:23:48,450 --> 00:23:53,399
the power consumption you could run it
in as part of a mobile lab if you're

258
00:23:53,399 --> 00:23:57,690
looking for phalates in the field for
example and essentially as I mentioned

259
00:23:57,690 --> 00:24:03,330
before the otm technology what it does
is is essentially by heated oven so

260
00:24:03,330 --> 00:24:06,570
essentially your capillary column is
inductively coupled with a heater that

261
00:24:06,570 --> 00:24:11,399
allows the maximum a lot of heat
transfer very efficient efficiently so

262
00:24:11,399 --> 00:24:15,749
now you have faster heating and cooling
times so your chromatography males will

263
00:24:15,749 --> 00:24:19,559
also improve you have more sample
throughput now chromatograms that will

264
00:24:19,559 --> 00:24:24,539
take you 20 to 25 minutes for analysis
may take you five or less minutes for

265
00:24:24,539 --> 00:24:26,530
the same
analysis and still having the same

266
00:24:26,530 --> 00:24:30,790
resolution and sensitivity so again
that's that's something for you if

267
00:24:30,790 --> 00:24:33,040
you're not aware of for you to think
about if you're thinking about doing

268
00:24:33,040 --> 00:24:42,730
mobile lab applications of phthalates in
the field okay so one of the things that

269
00:24:42,730 --> 00:24:46,510
they imagined that we that we recently
were excited about is the introduction

270
00:24:46,510 --> 00:24:53,110
of ion trap technology for GCMs and one
of the benefits that you receive with

271
00:24:53,110 --> 00:24:57,010
ion trap technologies the ability to do
multiple I announced a ssin techniques

272
00:24:57,010 --> 00:25:00,790
within the same chromatogram so for
example for an ion for particular

273
00:25:00,790 --> 00:25:06,670
phthalates that may not be as responsive
to e.i analysis you could do a CI and

274
00:25:06,670 --> 00:25:09,730
Alice chemical ionic station on the same
compound in the same run in same

275
00:25:09,730 --> 00:25:14,170
chromatogram so you can switch from e-i
to CI without changing the source in the

276
00:25:14,170 --> 00:25:19,720
same run which will allow you greater
sensitivity and specificity for analytes

277
00:25:19,720 --> 00:25:23,850
of interest and I'll show a particular
example of that in this case we have a

278
00:25:23,850 --> 00:25:32,500
VI MP and the IPP the IDP and we have
them in AI and CI if you notice the e I

279
00:25:32,500 --> 00:25:39,730
expect from shows the classical Mass 149
s as the after parent peak however if we

280
00:25:39,730 --> 00:25:46,840
use CI with MTBE liquid MTBE were able
to differentiate the two now base you

281
00:25:46,840 --> 00:25:53,770
see have the molecular ion be the parent
peak and now we have 419 and for 47 okay

282
00:25:53,770 --> 00:25:57,580
the molecular ions in a higher intensity
so if this is something that there may

283
00:25:57,580 --> 00:26:01,870
be interest of you for you to identify
the two you have the flexibility of

284
00:26:01,870 --> 00:26:08,050
doing CI a liquid CI with MTBE to allow
you to differentiate those two and give

285
00:26:08,050 --> 00:26:12,220
you that much more degrees of confidence
okay and again this is coupled to the

286
00:26:12,220 --> 00:26:17,380
same 7890 GCC pneumatics are you used to
on the front end just a different

287
00:26:17,380 --> 00:26:21,970
ionization technique iron trap
technology on the back end okay so here

288
00:26:21,970 --> 00:26:31,060
for example in terms of calibration
curve for bbp and VBP able to go down to

289
00:26:31,060 --> 00:26:38,640
levels of one parts per billion or less
okay so now we mentioned the other

290
00:26:38,640 --> 00:26:42,420
labor of mass spectrometer that can be
applied to the fellows analysis involves

291
00:26:42,420 --> 00:26:46,620
the use of a triple quadrupole GCMs and
I go briefly into how that operates

292
00:26:46,620 --> 00:26:51,360
currently in our lab in and Santa Clara
are one of our applications our chemists

293
00:26:51,360 --> 00:26:56,070
and in conjunction with dois we're
frontier we have a paralyzer system and

294
00:26:56,070 --> 00:27:02,070
you might hear some of that today but
essentially we're analyzing phthalates

295
00:27:02,070 --> 00:27:06,180
using triple quadruple GCMs and that
gives us a different degrees of

296
00:27:06,180 --> 00:27:09,660
confidence and also lower sensitivity
and there's an application now coming

297
00:27:09,660 --> 00:27:13,020
soon with that information but I talked
briefly of some of the preliminary

298
00:27:13,020 --> 00:27:17,730
results that we that we have obtained so
let me give you a little brief overview

299
00:27:17,730 --> 00:27:22,370
of how triple quadrupole GCMs operates
from for example you have your typical

300
00:27:22,370 --> 00:27:27,750
spectrum so you ionize your samples they
give you any I mode in this particular

301
00:27:27,750 --> 00:27:31,800
example you get a classical spectrum
okay everything is ionized broken down

302
00:27:31,800 --> 00:27:39,720
into individual components however
through the first quadrupole we isolate

303
00:27:39,720 --> 00:27:44,640
one ion in this case we select the most
intense ion which is 210 we'll select

304
00:27:44,640 --> 00:27:50,250
that ion in this particular case and
your Simenon die on that iron you pass

305
00:27:50,250 --> 00:27:53,460
it that iron through a collision cell
hit it with some energy and some

306
00:27:53,460 --> 00:27:58,380
collision gases you're able to break
down that ion even further into two

307
00:27:58,380 --> 00:28:02,280
individual components and then you can
monitor multiple ions from that

308
00:28:02,280 --> 00:28:06,990
secondary fragmentation in this case
will be a transition of for example to

309
00:28:06,990 --> 00:28:12,480
10 go into 158 and 191 which may give
you that much more fragmentation and

310
00:28:12,480 --> 00:28:17,760
identification for some of the trouble
and lights of interest okay so in this

311
00:28:17,760 --> 00:28:22,680
case we have a mrm chink thick total
iron chromatogram in of phthalates and

312
00:28:22,680 --> 00:28:30,630
lemon oil okay this is this is the mrm
take the e i spectra of the NOP and the

313
00:28:30,630 --> 00:28:35,550
triple quad and you can see that the
iron ratios are very similar to your

314
00:28:35,550 --> 00:28:41,070
nest libraries that you're used to and
we have developed transition we have

315
00:28:41,070 --> 00:28:44,880
developed mrm translation multiple
reaction monitored transaction for our

316
00:28:44,880 --> 00:28:49,560
different values of interest where now
if you're interested in having more

317
00:28:49,560 --> 00:28:54,809
information on your salads and
in that one particular ion with some

318
00:28:54,809 --> 00:29:00,450
energy and a collision flow you're able
to get to individual transitions one for

319
00:29:00,450 --> 00:29:05,490
quantitation and one for quality ssin
and you're able to develop a set up find

320
00:29:05,490 --> 00:29:08,970
ratios for each particular transition
therefore the ohmmeter particular

321
00:29:08,970 --> 00:29:13,080
requirements for your quantitation this
decreases your for your false positives

322
00:29:13,080 --> 00:29:15,840
was met what it's talking about about
false positives being a big problem

323
00:29:15,840 --> 00:29:18,600
ms/ms
could potentially solve that for you

324
00:29:18,600 --> 00:29:26,130
reduces the amount of false positives in
your thallus analysis okay now for for

325
00:29:26,130 --> 00:29:30,990
higher quality data and for more
interesting results for example you

326
00:29:30,990 --> 00:29:36,720
could take the power of Q Tov mass
accuracy and high resolution for the

327
00:29:36,720 --> 00:29:41,640
analysis of valleys in high matrix you
have the ability to do ms/ms and E I as

328
00:29:41,640 --> 00:29:49,440
well and in this case what we have is
our traditional triple quad which is

329
00:29:49,440 --> 00:29:53,970
based on your single quad mass
spectrometer that you used to currently

330
00:29:53,970 --> 00:29:57,049
use it in your lab we took that
technology a step further and

331
00:29:57,049 --> 00:30:01,409
incorporated our trouble quadrant mass
spectrometer to our tough mass

332
00:30:01,409 --> 00:30:05,580
spectrometer that we have in our we had
in our LC world that we made this

333
00:30:05,580 --> 00:30:09,330
marriage of a cute off so now you have
the ability of doing full scan high

334
00:30:09,330 --> 00:30:15,690
resolution high mass accuracy less than
5 ppm mass accuracy and you know I'll

335
00:30:15,690 --> 00:30:18,900
show you what that gives you what level
of information that provides you in a

336
00:30:18,900 --> 00:30:24,780
second so for example if you're used to
looking at your typical single quad mass

337
00:30:24,780 --> 00:30:29,970
spectrometer and we'll look at Matt's
mass 604 what you see in the red is your

338
00:30:29,970 --> 00:30:35,400
typical spectrum however it with the
tough technology you're able to really

339
00:30:35,400 --> 00:30:41,460
resolve the 613 and the 614 the 649 and
you get accurate masses to the point

340
00:30:41,460 --> 00:30:44,820
that you have more resolution and when
we talk about resolution we're talking

341
00:30:44,820 --> 00:30:49,230
about the space between the two the two
different isotopes there let you see

342
00:30:49,230 --> 00:30:53,340
okay so that's one level of confidence
that will also decrease your false

343
00:30:53,340 --> 00:30:56,100
positive or negative so if you have a
high matrix and things that may be

344
00:30:56,100 --> 00:31:01,260
interfering q top technology with the
high- accuracy and resolution may give

345
00:31:01,260 --> 00:31:04,840
you that extra level of confidence of
able to seize the needle in the haystack

346
00:31:04,840 --> 00:31:12,850
a little bit better okay so for example
your typical mass spectrometer has a has

347
00:31:12,850 --> 00:31:20,920
a insurgency let's say 0.3 a.m. use and
for mass 271 for 9:00 8:00 a.m. you

348
00:31:20,920 --> 00:31:26,740
there are 7600 compounds that are
possible based on that mass insurgency

349
00:31:26,740 --> 00:31:33,250
of 1000 ppm as the as the young
certainty decreases let's say to a level

350
00:31:33,250 --> 00:31:39,910
of 1 ppm parts per million now you have
11 possible compounds that are possible

351
00:31:39,910 --> 00:31:45,280
with that particular ion of interest so
that's where the again reduces your

352
00:31:45,280 --> 00:31:51,520
false positives or high level of degree
of confidence in your analysis also too

353
00:31:51,520 --> 00:31:57,220
so those of you that are using our our
mass spectrometers for the analysis of

354
00:31:57,220 --> 00:32:01,720
phthalates retention time locking is
very important and as Matt discussed

355
00:32:01,720 --> 00:32:05,830
earlier retention time is one of the
requirements for for the methods of

356
00:32:05,830 --> 00:32:11,830
analysis having retention time locking
allows you to have one method on one

357
00:32:11,830 --> 00:32:15,580
mass spectrometer transfer that method
to a second mass spectrometer and

358
00:32:15,580 --> 00:32:18,790
they'll have the exact retention time
okay

359
00:32:18,790 --> 00:32:22,300
and whether you perform current
maintenance or not you're able to just

360
00:32:22,300 --> 00:32:26,170
we lock the method it's very simple to
do it just takes five injection you

361
00:32:26,170 --> 00:32:29,230
could do it overnight or over a lunch
break

362
00:32:29,230 --> 00:32:33,970
or right you can do it actually while
you are commissioning your instruments

363
00:32:33,970 --> 00:32:38,290
for analysis doing your performance
check you take one analyte of interest

364
00:32:38,290 --> 00:32:43,150
you look for that ion what it does is
retention time locking what it does it

365
00:32:43,150 --> 00:32:51,070
incorporates all the possible variations
in column length pressure or whether you

366
00:32:51,070 --> 00:32:55,270
have a difference of a few millimeters
going into the inlet or not it takes all

367
00:32:55,270 --> 00:32:59,559
the possible errors incorporates them
into a pressure error and allows you to

368
00:32:59,559 --> 00:33:04,030
match your retention times with
fermented transfer okay so in any system

369
00:33:04,030 --> 00:33:07,360
across this so you have a lab and the
East Coast the west coast the retention

370
00:33:07,360 --> 00:33:12,670
times will be the same if we're trying
to match the new phthalates analytes of

371
00:33:12,670 --> 00:33:14,710
interest so that's yes you have a
question

372
00:33:14,710 --> 00:33:18,510
that within the same historian if you
were to replace a Kabul say could you

373
00:33:18,510 --> 00:33:28,180
pull the data up and then switch out the
column and then run that software runs

374
00:33:28,180 --> 00:33:33,539
and then have
yes yes you can do that while you change

375
00:33:33,539 --> 00:33:38,610
count that retention time if you want
the IP to be at 11 minutes 11 point oo 1

376
00:33:38,610 --> 00:33:41,010
minute
it will be at 11.1 minutes in this

377
00:33:41,010 --> 00:33:43,760
instrument and that instrument will you
change your column every day all day

378
00:33:43,760 --> 00:33:47,760
year after year and it's free is
included in your mouth and your camp

379
00:33:47,760 --> 00:33:51,690
station software it's a button if you go
in your method there will be an option

380
00:33:51,690 --> 00:33:56,910
instrument retention time lock and I'll
ask you to put a violin in a particular

381
00:33:56,910 --> 00:34:00,870
position position one it'll do five runs
at five different pressures and they

382
00:34:00,870 --> 00:34:04,230
don't make a calibration of pressure
versus retention time and then will

383
00:34:04,230 --> 00:34:08,220
retain that calibration file and then
you're able to your mentor will be

384
00:34:08,220 --> 00:34:11,429
locked to a particular compound so if
you want to lock whether on your

385
00:34:11,429 --> 00:34:15,540
internal standard or whatever it is you
can lock in your internal standard and

386
00:34:15,540 --> 00:34:19,350
there was those retention times will
match that way you won't have to

387
00:34:19,350 --> 00:34:23,159
question it this style aid or is it that
valid no because it's retention times

388
00:34:23,159 --> 00:34:29,820
will match all day every day okay again
and retention time lock and it's

389
00:34:29,820 --> 00:34:35,730
available everything from from our 7059
75e single quad to our cubes off okay

390
00:34:35,730 --> 00:34:39,720
even to your mobile lab if you want to
retention time in your mobile lab you

391
00:34:39,720 --> 00:34:45,899
have that flexibility so again one of
things I want to emphasize this and the

392
00:34:45,899 --> 00:34:49,500
other speakers will go into different
other front ends that that made you an

393
00:34:49,500 --> 00:34:52,619
analysis but I wanted to focus in the
different capabilities is the take-home

394
00:34:52,619 --> 00:34:56,550
message here is that if you're looking
for routine analysis and you want to do

395
00:34:56,550 --> 00:35:00,600
next and this is the convolution our
single quad will be do a great job if

396
00:35:00,600 --> 00:35:07,500
you haven't promised with hi matrix
interference then there's that's when a

397
00:35:07,500 --> 00:35:12,480
triple quad may be more suitable for the
analysis because we'll give you that mrm

398
00:35:12,480 --> 00:35:15,359
transition of one iron going through
different transitions with two different

399
00:35:15,359 --> 00:35:19,710
iron ratios so you get more degrees of
confidence and if going from E AI to CI

400
00:35:19,710 --> 00:35:22,980
is something that you really want to do
for your analysis to give you that peace

401
00:35:22,980 --> 00:35:26,070
of mind and give you that much more
confidence in your results that's when

402
00:35:26,070 --> 00:35:32,430
an iron trap with like liquid CI what
MTBE may may aid you even more I just

403
00:35:32,430 --> 00:35:36,510
wanna the references up you can look
these up in our application notes and

404
00:35:36,510 --> 00:35:39,210
there's more to come
especially specifically on the cubes off

405
00:35:39,210 --> 00:35:43,530
and the triple quad analysis and
I want to give a special thanks to our

406
00:35:43,530 --> 00:35:47,490
scientists that have worked on this
Stefan Bowman fred fighter herm robert

407
00:35:47,490 --> 00:35:50,700
kubis and matt thank you for inviting us
to be here today

408
00:35:50,700 --> 00:35:55,670
any questions yes

409
00:36:17,650 --> 00:36:24,980
yes oh yes the question was whether we
could use a ammonia CI reagent ah for

410
00:36:24,980 --> 00:36:28,280
the analysis of phthalates and the
answer is yes you can in this particular

411
00:36:28,280 --> 00:36:32,240
case I was MTBE was used for through the
analysis but you could definitely use

412
00:36:32,240 --> 00:36:35,540
ammonia or even methane of that which
but ammonia temps tends to be more

413
00:36:35,540 --> 00:36:40,100
sensitive in terms of CI from what we've
seen in particular for phthalates yes

414
00:36:40,100 --> 00:36:44,920
good question any any other questions
yes

415
00:36:45,580 --> 00:36:51,110
thus we tension time lock work for
multiple analytes the answer is yes what

416
00:36:51,110 --> 00:36:55,280
we tension time locking does is it takes
one particular analyte and typically in

417
00:36:55,280 --> 00:36:59,390
the middle of your chromatogram so you
pick whatever analytes and your lock on

418
00:36:59,390 --> 00:37:03,410
that and everything else is gonna be
relative to that so if you if this

419
00:37:03,410 --> 00:37:06,980
analyte match everything else should
match depending then you have a good

420
00:37:06,980 --> 00:37:10,790
column in your column is now bad so yes
so all the components will be retention

421
00:37:10,790 --> 00:37:16,060
time locked but you only like a want to
compound everything else is relative

422
00:37:27,520 --> 00:37:31,820
what broader Peaks retention time
locking what is gonna do is you're still

423
00:37:31,820 --> 00:37:36,050
gonna have that broad peak however that
broad peak will be at the same time so

424
00:37:36,050 --> 00:37:38,360
that doesn't say essentially what's
gonna happen it's gonna add your your

425
00:37:38,360 --> 00:37:41,270
your chromatography in terms of
resolution but it will give you that

426
00:37:41,270 --> 00:37:46,130
same time for that blob of gram that you
have for the InP yes good question yes

427
00:37:46,130 --> 00:37:48,760
another question

428
00:38:00,560 --> 00:38:05,580
okay the question is is our triple
Quadra trapping or ions and the answer

429
00:38:05,580 --> 00:38:09,840
is it's not really trap it is scanning
scanning very fast and it's taking a

430
00:38:09,840 --> 00:38:14,490
selectively passing ions that we select
that we want so for example in this

431
00:38:14,490 --> 00:38:20,040
particular example that I showed let's
say you want to isolate 149 I've won a

432
00:38:20,040 --> 00:38:24,810
and 149 will go through the quad it'll
reach a collision cell once it reaches

433
00:38:24,810 --> 00:38:28,710
the collision cell it will be hit with a
particular energy typically between ten

434
00:38:28,710 --> 00:38:32,820
and fifteen volts electron volt and a
collision gas and from there we'll have

435
00:38:32,820 --> 00:38:35,970
two different transitions that we can
monitor typically the most the two

436
00:38:35,970 --> 00:38:41,150
strongest transitions to give you that
much more of degrees of confidence

437
00:38:50,720 --> 00:38:54,410
the question was where is there any
gain-of-function automatic gain function

438
00:38:54,410 --> 00:38:58,040
the answer is there is no automatic gain
however for compounds of interest you

439
00:38:58,040 --> 00:39:02,000
could adjust the gains in the different
time segments so for example for

440
00:39:02,000 --> 00:39:05,750
analytes that are maybe less sensitive
you may adjust the gain at that

441
00:39:05,750 --> 00:39:19,369
particular moment in time thank you so
much up next we have Bob Freeman from

442
00:39:19,369 --> 00:39:34,760
frontier labs all right good morning I'd
like to thank Matt for the opportunity

443
00:39:34,760 --> 00:39:38,030
to talk about the method it's been in
development now for about two or three

444
00:39:38,030 --> 00:39:44,060
years about about three years ago we
started at the ASTM D 20 former

445
00:39:44,060 --> 00:39:50,359
committee to look at low-level
phthalates in polyvinyl chloride and for

446
00:39:50,359 --> 00:39:53,480
the past two or three years this is what
we focused on we've done two sets of

447
00:39:53,480 --> 00:39:58,280
round-robin samples and we've done
independent validation study with some

448
00:39:58,280 --> 00:40:01,730
laboratories so I'd like to give you
just a status report on where that

449
00:40:01,730 --> 00:40:05,930
method is give you a little bit of data
and then I'll also like to kind of

450
00:40:05,930 --> 00:40:10,369
address the issue of phthalates in hold
because you'll notice that for those

451
00:40:10,369 --> 00:40:14,660
people who do work on methods phthalates
our particular problem because the fact

452
00:40:14,660 --> 00:40:18,800
is that the method no longer is what
what it was designed to do we started

453
00:40:18,800 --> 00:40:23,540
off with toys and we started off with
six phthalates and in the last two or

454
00:40:23,540 --> 00:40:26,990
three years now all of a sudden we're
looking at electronic products we're

455
00:40:26,990 --> 00:40:32,420
looking at food containers we're looking
at a whole variety of different matrices

456
00:40:32,420 --> 00:40:36,440
and you'll notice that we've gone from
six phthalates till now there's interest

457
00:40:36,440 --> 00:40:40,609
in more than six in some cases fifteen
and sixteen phthalates so whereas we

458
00:40:40,609 --> 00:40:45,260
started with toys and consumer products
now we have to work the stakeholders

459
00:40:45,260 --> 00:40:51,589
include formulators past recyclers
manufacturers retailers consumers all of

460
00:40:51,589 --> 00:40:54,829
a sudden everybody's interested in
phthalates and what we find out is

461
00:40:54,829 --> 00:40:58,670
interesting that lights all over the
world and this is all well and good

462
00:40:58,670 --> 00:41:03,140
except for now you notice that the list
is starting to expand a little bit and

463
00:41:03,140 --> 00:41:06,530
so some of the electronics
News have included things like dye

464
00:41:06,530 --> 00:41:10,520
isobutyl phthalates
which is a which is a plasticizer that's

465
00:41:10,520 --> 00:41:16,400
not too not too uncommon it's a low-cost
replacement for DBP so now we've gone

466
00:41:16,400 --> 00:41:22,579
from from three regulated phthalates to
a possible six phthalates seventh a

467
00:41:22,579 --> 00:41:27,109
lates and you can see we're having the
typical encounter here and that is as we

468
00:41:27,109 --> 00:41:31,010
develop the method we officer may
include more and more matrices and as we

469
00:41:31,010 --> 00:41:34,069
do that we want more and more compounds
and as we do that of course we want more

470
00:41:34,069 --> 00:41:38,809
and more sensitivity and so those of us
are workout methods this is a continuing

471
00:41:38,809 --> 00:41:43,490
problem there's even there's even
individual countries that are writing

472
00:41:43,490 --> 00:41:48,740
general methods that have promulgated
these most of them in 2011 and these are

473
00:41:48,740 --> 00:41:52,460
just mentioning things like all products
containing phthalates I mean that's a

474
00:41:52,460 --> 00:41:59,660
pretty general description for method to
address and because now you hear more

475
00:41:59,660 --> 00:42:02,799
and more about the countries not being
important the global global

476
00:42:02,799 --> 00:42:06,770
international corporations really being
the boundaries there are many

477
00:42:06,770 --> 00:42:11,869
corporations that work on a global scale
and have their own in-house methods and

478
00:42:11,869 --> 00:42:15,619
in-house list of compounds and in
Houston Texas and limits for various

479
00:42:15,619 --> 00:42:18,920
number of phthalates and so these are
just some of the electronics companies

480
00:42:18,920 --> 00:42:23,990
that we know I'm familiar with that are
working on phthalates in recycled

481
00:42:23,990 --> 00:42:29,540
plastics and different lists and we work
on right now so we've gone from in the

482
00:42:29,540 --> 00:42:35,030
beginning the ASTM method is PVC and six
compounds so now we're talking about

483
00:42:35,030 --> 00:42:40,400
maybe fifteen or sixteen and way beyond
PVC so let me back up a little bit talk

484
00:42:40,400 --> 00:42:43,880
about the ASTM methods
we're looking for a standard practice

485
00:42:43,880 --> 00:42:47,900
for the determination of low level
phthalates and polyvinyl chloride and

486
00:42:47,900 --> 00:42:52,369
the method that seems to give us the
best results this thermal desorption GC

487
00:42:52,369 --> 00:42:56,750
mass spec so there's really five steps
in this method and this is the most

488
00:42:56,750 --> 00:43:00,530
difficult step to step right here called
sample homogeneity because when you take

489
00:43:00,530 --> 00:43:04,670
a toy for example different parts of
that toy are made of four different

490
00:43:04,670 --> 00:43:08,510
plastics and each of those different
parts of plastics can have a different

491
00:43:08,510 --> 00:43:12,049
level of phthalates and so I would say
if I if I were to look at the overall

492
00:43:12,049 --> 00:43:17,150
scope of analyzing for phthalates in
materials the biggest problem is a

493
00:43:17,150 --> 00:43:21,290
homogeneous
because if I take this toy apart I can

494
00:43:21,290 --> 00:43:23,360
get three different numbers some of them
are good and some of them are a little

495
00:43:23,360 --> 00:43:27,380
bit higher a circuit board forces the
same problem a food container proposes

496
00:43:27,380 --> 00:43:30,950
the same problem and so on and so forth
so this is the single most difficult

497
00:43:30,950 --> 00:43:36,410
thing to do and that is to get a
homogeneous sample that sounds nice to

498
00:43:36,410 --> 00:43:39,980
say crowd milling you go into
laboratories that do crowd milling and

499
00:43:39,980 --> 00:43:44,930
contamination crosstalk between Crowell
Mills is ever-present and so one of the

500
00:43:44,930 --> 00:43:48,860
problems with phthalates as we as we
mentioned here is contamination well

501
00:43:48,860 --> 00:43:53,300
right off the bat if you're using one
crowd milled into five toys the

502
00:43:53,300 --> 00:43:59,960
possibility for cross-contamination is
very high this is the toughest step but

503
00:43:59,960 --> 00:44:04,700
the method that the the thumb of the
ASTM method for based on thermal

504
00:44:04,700 --> 00:44:08,240
desorption GC mass spec uses two
solutions one is just a standard

505
00:44:08,240 --> 00:44:12,320
solution of the phthalates of interest
and the other is the sample itself which

506
00:44:12,320 --> 00:44:16,880
is dissolved in tetrahydrofuran so this
is the there's only two solutions

507
00:44:16,880 --> 00:44:22,490
involved the injection is simply placing
these solutions in a small container

508
00:44:22,490 --> 00:44:27,020
raising the temperature the containers
such that the target compounds evolve or

509
00:44:27,020 --> 00:44:31,250
make out or externally extracted and
analyzing just that portion of the

510
00:44:31,250 --> 00:44:36,500
sample that contains the phthalates
separation is done by a standard 30

511
00:44:36,500 --> 00:44:41,780
meter db5 or 5% phenol column and
detector you can pick up most of the

512
00:44:41,780 --> 00:44:46,100
work we've been done has been done with
scanning ms several of the validation

513
00:44:46,100 --> 00:44:51,170
labs used si MMS you can use quadruple
crippled quads you can use negative ion

514
00:44:51,170 --> 00:44:57,800
CI this is the area where the method
seems to be the most applicable and that

515
00:44:57,800 --> 00:45:02,240
is this is this is unlike the first talk
this is how to get the sampler prepared

516
00:45:02,240 --> 00:45:08,630
and how to get the sample into the
column okay what column you use is up to

517
00:45:08,630 --> 00:45:13,090
you it's independent of the column
choice what sort of detection algorithm

518
00:45:13,090 --> 00:45:19,400
system you want to use this up to you so
this is the sample prep the whole idea

519
00:45:19,400 --> 00:45:21,830
here is to give her the solvent to get
rid of the glassware

520
00:45:21,830 --> 00:45:25,220
because I've gone into laboratories and
who do this and they'll tell me

521
00:45:25,220 --> 00:45:30,020
oftentimes they spend more time cleaning
Sox lists and using Sox's your salads

522
00:45:30,020 --> 00:45:33,559
are difficult to get rid of
and so what we do here is we start off

523
00:45:33,559 --> 00:45:36,740
with the product some sort of material
could be a circuit board could be a duct

524
00:45:36,740 --> 00:45:41,240
we all like ducts you Crowl millet you
want the smallest pieces you can if

525
00:45:41,240 --> 00:45:46,309
possible you want dust because you want
a homogeneous bottle of dust you take a

526
00:45:46,309 --> 00:45:50,210
small portion of that and put it in a
volumetric as a ten milliliter

527
00:45:50,210 --> 00:45:55,609
volumetric so now you've got so much
weight you add to that THF you make a

528
00:45:55,609 --> 00:46:01,849
solution now so now you have a solution
a THF solution of your sample the method

529
00:46:01,849 --> 00:46:05,960
calls for taking ten microliters of that
put it in a small cup and evaporating a

530
00:46:05,960 --> 00:46:09,859
solvent such that you win with a very
thin film of the sample on the inside

531
00:46:09,859 --> 00:46:14,240
surface of the cup at the bottom so to
prepare a sample like this once the

532
00:46:14,240 --> 00:46:17,720
crowd millings been done to prepare a
sample like this takes on the order of

533
00:46:17,720 --> 00:46:21,290
five minutes
they only solve what you're going to use

534
00:46:21,290 --> 00:46:25,700
as 10 milliliters of the THF you're
going to use disposable glassware so

535
00:46:25,700 --> 00:46:29,660
there's no worry about contamination or
cross-contamination that little cup then

536
00:46:29,660 --> 00:46:34,040
is ready to be analyzed now if you
wanted to do a screening method it's not

537
00:46:34,040 --> 00:46:37,579
necessary to go through the thin film
process and that is you can take this

538
00:46:37,579 --> 00:46:41,990
sample dust from your jar and just weigh
in a hundred micrograms or 200

539
00:46:41,990 --> 00:46:46,520
micrograms and put it directly in the
cup so you analyze it as a solid if

540
00:46:46,520 --> 00:46:49,250
you'd like that internal standard at
that point right now just a monitor

541
00:46:49,250 --> 00:46:53,150
performance that's easily done so either
way you end up with the cup that's ready

542
00:46:53,150 --> 00:46:57,349
to analyze now the difference between a
thin film that just making a THF

543
00:46:57,349 --> 00:47:02,240
solution put it on the cup and then
evaporating the solvent and using the

544
00:47:02,240 --> 00:47:07,609
solid material is basically reflected in
the precision so if I use a THF solution

545
00:47:07,609 --> 00:47:13,309
this is a half a milligram of sample now
you can see for this de HB the percent

546
00:47:13,309 --> 00:47:18,500
relative standard deviation is about one
percent for injections if I use the

547
00:47:18,500 --> 00:47:22,369
solid form for a screening method where
I just take the dust if you will five

548
00:47:22,369 --> 00:47:26,240
point five milligrams of dust to put in
the cup and analyze it directly you can

549
00:47:26,240 --> 00:47:30,799
see my precision goes down and of course
the reason for that is is if I look at

550
00:47:30,799 --> 00:47:35,240
this this collections best that I have
the chances of having the same amount of

551
00:47:35,240 --> 00:47:40,040
particles in each sample remote so I
give it a more scatter than the data

552
00:47:40,040 --> 00:47:44,119
where if I make up a solution I can
analyze that same solution numerous

553
00:47:44,119 --> 00:47:48,590
times and get exactly the same
but for a screening technique that the

554
00:47:48,590 --> 00:47:53,990
solid analysis seems to work just fine
can we hope to the end here thank you

555
00:47:53,990 --> 00:47:59,930
so a sample assistant a little cup this
is on top of a liner - maple injection I

556
00:47:59,930 --> 00:48:05,450
dropped the cup in the furnace the
furnace is at 100 degrees I hate the

557
00:48:05,450 --> 00:48:11,630
furnace 200 320 degrees and as I heat it
up the phthalates will evolve and those

558
00:48:11,630 --> 00:48:15,950
are carried on to the column these
little cups are can replace a lot of

559
00:48:15,950 --> 00:48:20,420
stuff but I can do a number of samples
now one sample per hour the reason for

560
00:48:20,420 --> 00:48:23,750
that is is because the calibration the
quantitation is done by a standard

561
00:48:23,750 --> 00:48:29,030
addition that requires two analysis one
of the sample and one of the samples has

562
00:48:29,030 --> 00:48:34,400
been spiked with a known amount so
that's to analysis so by about one an

563
00:48:34,400 --> 00:48:38,570
hour if you're going to do a screening
sample that means one I don't have to

564
00:48:38,570 --> 00:48:42,110
program from one two three twenty I just
drop it into three twenty so I don't

565
00:48:42,110 --> 00:48:46,100
have that 11 minute to feel real sample
prep for extraction simple and then I

566
00:48:46,100 --> 00:48:52,180
could do about three per hour in the
screening mode so the sample is heated I

567
00:48:52,180 --> 00:48:56,450
thermally extract the phthalates of
interest they go on to the column and I

568
00:48:56,450 --> 00:49:02,510
get my chromatogram the conditions I'm
using for my survey motor basically 100

569
00:49:02,510 --> 00:49:08,570
to 700 at a fairly high rate that could
allow me to profile a characterize the

570
00:49:08,570 --> 00:49:12,080
sample and it tells me where the
phthalates come in the temperature zone

571
00:49:12,080 --> 00:49:15,770
of which the phthalates evolve and then
I do my thermal desorption my thermal

572
00:49:15,770 --> 00:49:21,710
desorption basically is 100 to 320 at 20
degrees a minute if you use in a thin

573
00:49:21,710 --> 00:49:25,250
film you can do it 40 degrees a minute
rigea moment much difference this goes

574
00:49:25,250 --> 00:49:31,430
on to a column at 80 degrees so here's
what it looks like the top traces by

575
00:49:31,430 --> 00:49:35,630
screen I don't have to do the screening
for every sample the importance of the

576
00:49:35,630 --> 00:49:40,100
screen is it characterizes the sample if
I'm doing PVCs forever I only have to do

577
00:49:40,100 --> 00:49:43,970
this once if I'm doing polypropylene I
only have to do it once but if I'm doing

578
00:49:43,970 --> 00:49:48,950
circuit boards or if I'm doing different
kinds of materials then maybe I have to

579
00:49:48,950 --> 00:49:53,930
do my screen to find out where my if my
phthalates have shifted but you can see

580
00:49:53,930 --> 00:49:58,970
if I'm the top trace this is this a PVC
sampled inch as a plasticizer I get a

581
00:49:58,970 --> 00:50:01,820
large group
of compounds out fester phthalates any

582
00:50:01,820 --> 00:50:06,040
dents may be present there's also a peak
for HCl because that's the decomposition

583
00:50:06,040 --> 00:50:11,450
product of polyvinyl chloride and then
of course I get my fractions out there

584
00:50:11,450 --> 00:50:18,980
right polymerized or I realize the PVC
so I find my zone by looking at ions and

585
00:50:18,980 --> 00:50:22,790
you can see the lower trace where it
says phthalates 149 you can see all the

586
00:50:22,790 --> 00:50:28,970
149 all my phthalates are out by 320 and
so my thermal desorption zone is 100 to

587
00:50:28,970 --> 00:50:33,440
320 at 20 degrees a minute it's a nice
thing about this is by doing this I'm

588
00:50:33,440 --> 00:50:36,950
going to extract those phthalates and
put those on the columns for separation

589
00:50:36,950 --> 00:50:42,910
but I leave behind in the cup all the
polymer so I don't have to worry about

590
00:50:42,910 --> 00:50:47,600
faking the pulp making the system system
at the cleanup the polymer I don't have

591
00:50:47,600 --> 00:50:51,680
to worry about contamination of my
system as I backflash or for the polymer

592
00:50:51,680 --> 00:50:54,770
just stays in the cup of when I'm done I
watched a polymer out and reuse the cup

593
00:50:54,770 --> 00:51:00,020
or get a new cup all I'm going to inject
under the GC is just that fracks of a

594
00:51:00,020 --> 00:51:04,570
sample that's extracted at 320 degrees

595
00:51:04,900 --> 00:51:09,920
so I can identify different compounds by
ions this right here again this is just

596
00:51:09,920 --> 00:51:13,640
your standard this is probably more
interesting right here this is the

597
00:51:13,640 --> 00:51:18,290
reproducibility this is 6 injections of
a PVC dense sample and it's a little

598
00:51:18,290 --> 00:51:22,370
hard to read but what I have here is
this percent relative standard deviation

599
00:51:22,370 --> 00:51:27,560
of the precision of a multiple runs of
the same sample and you can see the

600
00:51:27,560 --> 00:51:33,560
precision across the board is less than
less than 2% so I can take one solution

601
00:51:33,560 --> 00:51:38,530
and analyze the six times and my
precision is pretty darn good

602
00:51:39,460 --> 00:51:43,220
remember I mentioned that this is a
standard addition method so I'm going to

603
00:51:43,220 --> 00:51:47,690
run my sample and get an area number for
my compound and then I'm going to spike

604
00:51:47,690 --> 00:51:51,730
the sample with a known amount in this
case point one five percent

605
00:51:51,730 --> 00:51:57,200
concentration and going to rerun the
sample so this is just the spiked sample

606
00:51:57,200 --> 00:52:00,830
I just show three of these right here so
this is the PVC dense sample where

607
00:52:00,830 --> 00:52:04,910
that's been spiked with a known amount
of each of the phthalates and again you

608
00:52:04,910 --> 00:52:09,430
can see the precision across the bottom
of the spiked numbers again less than 2%

609
00:52:09,430 --> 00:52:14,150
so it's very precise
I end up with a series of curves like

610
00:52:14,150 --> 00:52:19,400
this so these are the six calibration
curves for the six compounds the one in

611
00:52:19,400 --> 00:52:23,870
the middle is that the zero that's the
sample number the one on the far right

612
00:52:23,870 --> 00:52:28,010
that's the spike number and there's a
meth Max or go through and I can

613
00:52:28,010 --> 00:52:34,850
calculate the amount of amount of pieces
of phthalates in the original sample and

614
00:52:34,850 --> 00:52:39,200
if you do that again you can see the
precision there is less than 2% and it's

615
00:52:39,200 --> 00:52:43,940
a little hard to read but the numbers
the concentration of this of each tally

616
00:52:43,940 --> 00:52:49,670
that this particular one was about 980
990 nanograms so the accuracy is pretty

617
00:52:49,670 --> 00:52:54,830
good now that you see it korea is also
interested in this method this is some

618
00:52:54,830 --> 00:52:58,250
of their work these are their six
calibration curves only in this case

619
00:52:58,250 --> 00:53:01,820
they spiked with three mount three
different levels and all you do is take

620
00:53:01,820 --> 00:53:05,660
that one solution and you can see from
your upper right trace one case you put

621
00:53:05,660 --> 00:53:08,780
two microliters the next case three
microliters the next case 5 microliters

622
00:53:08,780 --> 00:53:12,650
x8 merica there's that's how you get you
multiple points so it's still one

623
00:53:12,650 --> 00:53:17,150
solution and you can see their
recoveries are of all the six phthalates

624
00:53:17,150 --> 00:53:22,490
are pretty good and their precision is
also less than one one percent in this

625
00:53:22,490 --> 00:53:27,170
case now we didn't mess the salvation
study that is we've been out to found

626
00:53:27,170 --> 00:53:30,140
five laboratories who are willing to try
this these as laboratories that hadn't

627
00:53:30,140 --> 00:53:34,850
done this before in one case one of the
laboratories had a new higher to that to

628
00:53:34,850 --> 00:53:39,230
the study force and this is this is what
we found from the validation study

629
00:53:39,230 --> 00:53:44,090
I sent them four samples to them with
the same sample the thermal desorption

630
00:53:44,090 --> 00:53:49,340
temperature system dependent by that I
mean if you have a factory system or a

631
00:53:49,340 --> 00:53:55,040
Pvt in the system or a CVS system or a
Gerstel system whatever system you're

632
00:53:55,040 --> 00:53:59,240
using for your thermal desorption you
have to be very sure that what you're

633
00:53:59,240 --> 00:54:03,830
looking at is the sample temperature or
not the thermocouple temperature so

634
00:54:03,830 --> 00:54:06,800
there's some differences in some systems
between the set point and the actual

635
00:54:06,800 --> 00:54:10,730
point how you ready to do is run the the
EGA and you can determine the

636
00:54:10,730 --> 00:54:15,320
temperature range of a which that system
operates best some of the validation

637
00:54:15,320 --> 00:54:20,810
labs you sim some of the validation labs
you scan it both to produce comparable

638
00:54:20,810 --> 00:54:25,310
data thin photo columns are definitely
preferred over thick film columns we

639
00:54:25,310 --> 00:54:27,090
don't
a lot of retention and a whole lot of

640
00:54:27,090 --> 00:54:30,480
time in there the standard edition
everyone use standard edition at one

641
00:54:30,480 --> 00:54:34,980
point standard edition the r-squared was
always greater than 0.99 if you if you

642
00:54:34,980 --> 00:54:38,520
recalled back to this study this is this
is three points plus the sample in the R

643
00:54:38,520 --> 00:54:43,260
squares and all these are one but these
five laboratories all data point nine

644
00:54:43,260 --> 00:54:47,340
nine are better the inner lab injection
to Jackson reproducibility is less than

645
00:54:47,340 --> 00:54:51,450
five percent the data I've shown here is
all about two percent one of my five

646
00:54:51,450 --> 00:54:56,400
labs with the newbie they were they were
about thirty percent higher than

647
00:54:56,400 --> 00:54:59,700
everybody else in the whole wide world
so there was some systematic issue there

648
00:54:59,700 --> 00:55:03,990
but I included all their data that's how
it comes out to be five percent the

649
00:55:03,990 --> 00:55:08,070
inter lab sample the sample
reproducibility is always less than ten

650
00:55:08,070 --> 00:55:13,380
percent and the laboratory laboratory
accuracy is 25 percent now get if I take

651
00:55:13,380 --> 00:55:17,460
out the data from that one lab and drop
down to about ten percent but I can't

652
00:55:17,460 --> 00:55:20,880
find a reason to do that so these are
the numbers we came up with from the

653
00:55:20,880 --> 00:55:26,310
validation study of these five
laboratories now all of a sudden we're

654
00:55:26,310 --> 00:55:28,950
getting requests for an extended target
compound list so here's the extended

655
00:55:28,950 --> 00:55:32,820
target compound this we've been asked to
look at next is this method I've

656
00:55:32,820 --> 00:55:36,810
highlighted in red the six compounds we
started with in order to do this

657
00:55:36,810 --> 00:55:39,300
chromatic graphically it's a simple
thing he just dropped the initial

658
00:55:39,300 --> 00:55:42,720
temperature from 100 to 80 degrees and
he tried to do a little bit better

659
00:55:42,720 --> 00:55:47,370
separation on that front end better stow
your host that only goes up to D ID P so

660
00:55:47,370 --> 00:55:50,730
it doesn't go any further so the
analysis time is still on the order of

661
00:55:50,730 --> 00:55:54,660
about twenty minutes so it has already
lengthen your analysis time and since

662
00:55:54,660 --> 00:55:58,650
your sample prep time is baking up a ten
microliters 10 milliliter solution of

663
00:55:58,650 --> 00:56:01,890
THF it's it doesn't affect any of your
sample prep and your operational

664
00:56:01,890 --> 00:56:07,470
parameters so it looks that was like
it's almost too good to be true we've

665
00:56:07,470 --> 00:56:14,100
done this for three years the TD method
is easy it takes one syringe and a

666
00:56:14,100 --> 00:56:17,130
little bit of glassware it's clean
there's no cleaning the cleaning and

667
00:56:17,130 --> 00:56:20,550
cleaning and cleaning and socks with
extractors that sort of thing it's great

668
00:56:20,550 --> 00:56:25,290
because I'm using almost no solvent at
all and it's fully automated that sounds

669
00:56:25,290 --> 00:56:28,380
really really good it's precise and
accurate the precision is always less

670
00:56:28,380 --> 00:56:33,120
than 2% the accuracy is plus or minus
10% in your laboratory and it

671
00:56:33,120 --> 00:56:37,200
accommodates an extended come down list
so this is all well and good there is a

672
00:56:37,200 --> 00:56:41,400
downside though
the downside is this run to run

673
00:56:41,400 --> 00:56:48,329
contamination issue was a real problem
the and thousand for millions that not

674
00:56:48,329 --> 00:56:53,219
too much of an issue that's bucket
chemistry but as the as the detection as

675
00:56:53,219 --> 00:56:57,779
the sensitivity requirements get higher
and higher more and more sensitivity we

676
00:56:57,779 --> 00:57:01,289
have to wait find a way to really
monitor this contamination the

677
00:57:01,289 --> 00:57:04,019
contamination is in the injection port
there's no doubt about it

678
00:57:04,019 --> 00:57:09,239
also the quantiles would have to be
verified as we change matrices that's

679
00:57:09,239 --> 00:57:13,380
one thing to look at PVC which is what
the ASTM method was focused on and we

680
00:57:13,380 --> 00:57:16,109
know appropriate probably propylene and
probably either ethylene and we've done

681
00:57:16,109 --> 00:57:18,900
a number of different polymers what
happens with the circuit board what

682
00:57:18,900 --> 00:57:21,960
happens with it some kind of material
you haven't really looked at so you

683
00:57:21,960 --> 00:57:25,589
really have to verify that you have no
in no interference when you start

684
00:57:25,589 --> 00:57:29,219
changing matrices and the second the
third thing is when you deal with these

685
00:57:29,219 --> 00:57:34,619
compounds that are isomeric mixtures
like di DP the quantitation becomes a

686
00:57:34,619 --> 00:57:38,249
little bit more difficult it could be I
love my mother would kill me for saying

687
00:57:38,249 --> 00:57:42,509
this it could be HPLC may be a better
way to do these multiple these compounds

688
00:57:42,509 --> 00:57:46,170
that have a lot of different isomers so
in the end those of us that develop

689
00:57:46,170 --> 00:57:49,920
methods are in the deep trouble here
because now we've got them we've got a

690
00:57:49,920 --> 00:57:54,479
method now that was developed for six
compounds in PVC and we're asked to do

691
00:57:54,479 --> 00:57:58,319
more and more phthalates and we're asked
to do it more and more different

692
00:57:58,319 --> 00:58:03,359
matrices remember being asked get lower
and lower and higher

693
00:58:03,359 --> 00:58:07,499
I guess higher and higher sensitivity
and in the case of phthalates this could

694
00:58:07,499 --> 00:58:12,960
be a real challenge for the laboratory
any questions we have time for one or

695
00:58:12,960 --> 00:58:17,930
two questions yes

696
00:58:19,069 --> 00:58:28,640
that was our SD company yeah yeah that's
what a half safer yes

697
00:58:36,510 --> 00:58:41,349
yeah the question is is the precision
accuracy depended upon of the cryogenic

698
00:58:41,349 --> 00:58:47,920
billing of a sample all the work here
was done with PVC standards that were

699
00:58:47,920 --> 00:58:52,630
prepared by one of the members of the
ASTM committee and so it wasn't

700
00:58:52,630 --> 00:59:15,910
literally a sheet of PVC and so yes see
now you're getting into legal issues

701
00:59:15,910 --> 00:59:19,599
yeah I mean obviously you are as a
citizen I think you are but as the the

702
00:59:19,599 --> 00:59:24,089
methodology that we've familiar with it
calls for a number for the entire

703
00:59:24,089 --> 00:59:29,109
product and so yes you are diluting the
beak but on the on the other end you

704
00:59:29,109 --> 00:59:34,690
know how much the because of the duct
you know what but that's the critical

705
00:59:34,690 --> 00:59:38,640
issues at homogeneity sample homogeneity

706
00:59:40,890 --> 00:59:47,650
just quickly to add on to that that
we're really not here to talk about the

707
00:59:47,650 --> 00:59:54,910
policy and the regulations but we have
issued guidance that it's the component

708
00:59:54,910 --> 01:00:02,829
parts that the thallus limit applies to
so we're our very initial guidance was

709
01:00:02,829 --> 01:00:11,950
different based on the initial read of
the rule we have guided the individual

710
01:00:11,950 --> 01:00:19,270
component parts but regardless of that
then the analysis is still the same all

711
01:00:19,270 --> 01:00:21,390
right

712
01:00:25,349 --> 01:00:30,750
up next we have chrissteele from bureau
veritas

713
01:00:38,700 --> 01:00:42,839
you could just press forward or back or
you can press one of these buttons

714
01:00:42,839 --> 01:00:49,059
doesn't make sure you speak clear enough
came back alright thanks Matt yes I'm

715
01:00:49,059 --> 01:00:53,829
filling in for Lisa Clarissa a global
technical consultant specialist because

716
01:00:53,829 --> 01:00:58,599
she is there's a touch of a flu bug
going spreading around at our Buffalo

717
01:00:58,599 --> 01:01:03,460
office so hopefully I don't get it after
after we leave to get back to Buffalo

718
01:01:03,460 --> 01:01:12,150
but so I'm speaking on behalf of Jarrow
Veritas in lieu of her so yes no problem

719
01:01:12,150 --> 01:01:18,160
so thank you Matt for letting me
represent our company here and I'm here

720
01:01:18,160 --> 01:01:22,839
today to talk about using a dual
instrumentation for phthalates analysis

721
01:01:22,839 --> 01:01:29,559
and plastics and what I mean by that is
two systems to mass spectrometers the

722
01:01:29,559 --> 01:01:33,430
first of course GCMs has I think a lot
of you are familiar with and that the

723
01:01:33,430 --> 01:01:39,579
CPSC method revolves around but secondly
the lc-ms system using both in

724
01:01:39,579 --> 01:01:47,589
conjunction with each other just to kind
of rule out any interferences that GCMs

725
01:01:47,589 --> 01:01:52,680
may pose basically to get like a second
set of eyes looking at a particular

726
01:01:52,680 --> 01:01:58,630
sample scan bureau veritas his head
about I'm guessing at least 10 years of

727
01:01:58,630 --> 01:02:02,829
experience testing for phthalates in a
variety of consumer products many

728
01:02:02,829 --> 01:02:08,859
different plastics and I'm here to try
to share with you hopefully some help

729
01:02:08,859 --> 01:02:13,359
for helpful information that has worked
for us in our testing approach to

730
01:02:13,359 --> 01:02:18,819
resolve these tricky samples as we know
as time goes on and regulations are

731
01:02:18,819 --> 01:02:23,470
expanding manufacturers seem to be
trying to get more and more clever as

732
01:02:23,470 --> 01:02:29,770
trying to find workarounds to find
alternatives plasticizer alternatives to

733
01:02:29,770 --> 01:02:36,069
have their samples compliance so in
effect you're going to we see it we see

734
01:02:36,069 --> 01:02:40,869
a lot of different scans and our thalut
extracts because of this because they're

735
01:02:40,869 --> 01:02:43,230
using

736
01:02:43,289 --> 01:02:51,400
alternatives sure sure sorry so that's
that's my purpose here today is to try

737
01:02:51,400 --> 01:02:59,260
to share with you our daily testing
approach so just as a quick outline I'm

738
01:02:59,260 --> 01:03:05,890
gonna go over the test method that we
use the instrumentation that we use GCMs

739
01:03:05,890 --> 01:03:12,880
and LCS LCMS of course results in
interpretation and I'll answer try to

740
01:03:12,880 --> 01:03:21,670
answer any questions you may have about
our approach our test method extremely

741
01:03:21,670 --> 01:03:28,960
similar to what's in CPSC essentially we
cut the sample into very small pieces of

742
01:03:28,960 --> 01:03:34,140
the component we place it to a
disposable glass reaction vessel or vial

743
01:03:34,140 --> 01:03:42,099
then we add tetrahydrofuran THF and
sonicated at 40 degrees centigrade for

744
01:03:42,099 --> 01:03:47,079
30 minutes which in most cases will be
long enough and aggressive enough to get

745
01:03:47,079 --> 01:03:52,359
that into the organic solution in Sonic
8 an additional 30 minutes if that

746
01:03:52,359 --> 01:03:56,349
sample does not visibly dissolve
sometimes that happens with harder

747
01:03:56,349 --> 01:04:05,470
plastics after which acetyl nitrile is
what we use acetyl nitrile ACN is that a

748
01:04:05,470 --> 01:04:13,059
drop ice to precipitate off the polymer
that's been very useful for us I know

749
01:04:13,059 --> 01:04:18,069
other solvents and the CBC mentions
hexane using hexane

750
01:04:18,069 --> 01:04:24,849
other LEDs may be using methanol they
may work in particular applications or

751
01:04:24,849 --> 01:04:32,109
plastics PVC probably works but we've
found that co nitrile will precipitate

752
01:04:32,109 --> 01:04:39,609
out more polymer more types of polymers
than the other solvents so in in the end

753
01:04:39,609 --> 01:04:47,829
it results in a cleaner layer on top of
your precipitated plasticizer a very

754
01:04:47,829 --> 01:04:53,740
precipitated polymer so you're all that
junk and interfering things that may

755
01:04:53,740 --> 01:04:57,769
interfere what the analysis will come
out of that solution and then you can

756
01:04:57,769 --> 01:05:06,529
inject a much more clean solution into
your instrument so once that is done

757
01:05:06,529 --> 01:05:14,089
you'll have to stand for 30 minutes and
take that solution and to put it into an

758
01:05:14,089 --> 01:05:20,029
autosampler vial and it's ready for
analysis actually ad if they're seeing

759
01:05:20,029 --> 01:05:24,829
good rate of the unforeseen there's an
internal standard then you can filter it

760
01:05:24,829 --> 01:05:36,579
and analyze it by GCMs indoor lcms and
currently LCMS is not listed in the cps

761
01:05:36,579 --> 01:05:45,109
c method currently I will go most of you
probably are already aware of the

762
01:05:45,109 --> 01:05:51,199
function of gas chromatography you have
a carrier gas usually helium the sample

763
01:05:51,199 --> 01:05:57,529
is injected in the injector volatilized
goes to the column the analyte separate

764
01:05:57,529 --> 01:06:01,789
out from one another as well as any
interfering compounds they separate out

765
01:06:01,789 --> 01:06:06,739
from one another then it enters the mass
spec detector fragmentation occurs a

766
01:06:06,739 --> 01:06:15,559
mass spectrum is produced and the signal
is plotted on the chromatogram okay what

767
01:06:15,559 --> 01:06:19,669
we're looking at here is just an example
of many different types of phthalates

768
01:06:19,669 --> 01:06:24,769
that are out there obviously it's not a
complete list but it's just it's just

769
01:06:24,769 --> 01:06:29,049
here to demonstrate what we could
potentially be looking at when we

770
01:06:29,049 --> 01:06:35,929
analyze a particular sample extract any
one of these or a combination of these

771
01:06:35,929 --> 01:06:42,799
could be found in your sample so a lot
of them are nice clean needle peaks of

772
01:06:42,799 --> 01:06:46,909
course as mad as mention and many of you
already know we have the broad finger

773
01:06:46,909 --> 01:06:53,989
peak type sale lights to that a
regulated di MPD IDP you can't see it

774
01:06:53,989 --> 01:06:58,339
too much on this slide but you know it's
it's buried in there between the 12 and

775
01:06:58,339 --> 01:07:04,669
13 minute range so this is just to
demonstrate what we could be looking at

776
01:07:04,669 --> 01:07:07,460
in a sample and like I said this is not
a complete with

777
01:07:07,460 --> 01:07:18,320
by any means some of these are regulated
some are not so go on to our next slide

778
01:07:18,320 --> 01:07:26,180
this is just an example of a mass
spectrum of DBP just as an example we go

779
01:07:26,180 --> 01:07:31,220
back a slide DBP is just one that's
highlighted as an example if we look at

780
01:07:31,220 --> 01:07:36,050
the mass spectrum creating that response
this is it

781
01:07:36,050 --> 01:07:44,630
of course they speak 149 is very common
qualifying ion could be 223 which is

782
01:07:44,630 --> 01:07:53,390
unique to DBP now a GC mass spectrum of
di NPD IDP those troublesome finger

783
01:07:53,390 --> 01:08:00,500
peaks you could the InP would be 293 you
could extract that out of the

784
01:08:00,500 --> 01:08:05,480
chromatogram but even doing so it still
can sometimes be a challenge depending

785
01:08:05,480 --> 01:08:09,800
on what what you might have interfering
with that I'm going to talk a little bit

786
01:08:09,800 --> 01:08:17,540
more about that in the upcoming slides
as well this is an example of a non

787
01:08:17,540 --> 01:08:22,520
regulated daylight currently unregulated
you never know what the future holds dye

788
01:08:22,520 --> 01:08:28,640
ISO octal phthalates basically this is a
finger peak as well this may be present

789
01:08:28,640 --> 01:08:32,509
a sample you don't know what
manufacturers are putting in we've

790
01:08:32,509 --> 01:08:38,270
certainly seen it many times before in
samples dye ISO octal phthalates it's a

791
01:08:38,270 --> 01:08:43,580
it's like a couple covered carbon chains
last of course then Diet di n P and D

792
01:08:43,580 --> 01:08:49,790
IDP the nanos and decals so here you can
see their retention time it's a range

793
01:08:49,790 --> 01:08:56,630
it's a broad range it's it's in isomeric
mix of course and that's why it looks

794
01:08:56,630 --> 01:09:02,029
the way it does so you can see it spans
across the time line retention time line

795
01:09:02,029 --> 01:09:09,290
you know between like eleven point two
five minutes to twelve and a half so is

796
01:09:09,290 --> 01:09:14,270
if you see something like this in a
sample it can be challenging to properly

797
01:09:14,270 --> 01:09:20,540
integrate the regulated phthalates or
even see if they're there di NPD IDP

798
01:09:20,540 --> 01:09:25,520
even after ion extraction honing in on a
particular ion that may be unique to

799
01:09:25,520 --> 01:09:41,170
that particular Valley so here is an
example of di MP on a GC chromatogram

800
01:09:41,170 --> 01:09:46,460
highlighted there it's it's if you want
to call that in a pax it's peeking out

801
01:09:46,460 --> 01:09:51,350
around 13 minutes but the range of
isomers begins at about 12 and a half

802
01:09:51,350 --> 01:09:58,520
and extends out to like 13 and a half
minutes so anyone doing a lead analysis

803
01:09:58,520 --> 01:10:05,410
by GC this is second nature I'm sure see
this all the time so again it's very

804
01:10:05,410 --> 01:10:12,080
difficult to integrate and try to
accurately capture the area counts of

805
01:10:12,080 --> 01:10:21,170
that di e MP peak if you have other
interferences or other phalates present

806
01:10:21,170 --> 01:10:26,170
it just it can be challenging on the GC

807
01:10:31,500 --> 01:10:36,780
so here we're just looking at some
chemical structures of some regulated

808
01:10:36,780 --> 01:10:44,250
thelet's we have d NOP di n p di DP d
NOP is unbranched so that's it it's a

809
01:10:44,250 --> 01:10:53,100
single peak but all three of these have
roughly the same retention time on GC D

810
01:10:53,100 --> 01:10:58,920
NOP is an octal di o P which I mentioned
before is an ISO octal so you're going

811
01:10:58,920 --> 01:11:04,740
to see 279 ion 279 even if you extract
enough for DN o P if you have di o P in

812
01:11:04,740 --> 01:11:10,680
your sample and what do you do you see
this big under al unregulated 279 I in

813
01:11:10,680 --> 01:11:16,080
chromatogram and then you may or may not
see this little peak come out of da that

814
01:11:16,080 --> 01:11:21,030
may or may not be DN o P if it is DN o P
and you're confident it is okay now how

815
01:11:21,030 --> 01:11:24,900
do you quantico annotate it it's very
hard to baseline integrate when it's

816
01:11:24,900 --> 01:11:34,320
coming out of a mess
so di NP + di D P of course our isomeric

817
01:11:34,320 --> 01:11:41,790
finger Peaks they actually overlap on GC
even though they have quantifying their

818
01:11:41,790 --> 01:11:48,210
qualifying ions associated with them
they can be troublesome as well and of

819
01:11:48,210 --> 01:11:54,450
course the reason that they are broad
finger peaks is because I so know no di

820
01:11:54,450 --> 01:12:03,350
NP and I sewed a co di DP have it could
be a multiple of arrangements of the

821
01:12:03,350 --> 01:12:08,640
groups on methyl groups they could be
located in many different statistical

822
01:12:08,640 --> 01:12:13,110
possibility and that's why you see that
broad finger peak of course it's not

823
01:12:13,110 --> 01:12:17,340
just a single analyte it's a mixture of
isomers I'm sure you're all familiar

824
01:12:17,340 --> 01:12:22,130
with why they look the way they do on GC

825
01:12:22,190 --> 01:12:27,540
now let's get into
lcms instrumentation liquid

826
01:12:27,540 --> 01:12:33,210
chromatography coupled with a mass
spectrometer pretty much basically the

827
01:12:33,210 --> 01:12:38,550
same principles GCMs you're just using
liquid is the mobile phase or carrier to

828
01:12:38,550 --> 01:12:45,070
carry along your analytes they get
separated by a column the same

829
01:12:45,070 --> 01:12:52,000
renters spray chamber mobile phases
evaporated fragmentation occurs and then

830
01:12:52,000 --> 01:12:58,230
the resulting ions are separated out by
mass to charge ratio and response is

831
01:12:58,230 --> 01:13:09,130
created as well as a spectrum so this is
an example this is a di NP at 1 ppm on

832
01:13:09,130 --> 01:13:16,270
the LCMS so what we're looking at here
is we run in scan mode this is actually

833
01:13:16,270 --> 01:13:23,199
I in 419 I believe di NP and this is
only 1 ppm we do run standards a little

834
01:13:23,199 --> 01:13:29,050
bit lower than this but the point is you
can see how much better it looks to a

835
01:13:29,050 --> 01:13:33,520
chemist I mean this this I like the way
it looks it's very integrated the

836
01:13:33,520 --> 01:13:42,460
signal-to-noise ratio is very large it's
just just a great responding signal for

837
01:13:42,460 --> 01:13:47,679
VI NP it just jumps out of the baseline
is it's it's great

838
01:13:47,679 --> 01:13:57,329
di DP at 1 ppm similar similar look to
it comes right out of the baseline you

839
01:13:57,329 --> 01:14:02,440
know you get a much better response of
course than compared to GC there's no

840
01:14:02,440 --> 01:14:10,409
broad finger peaks and actually the IDP
is fully resolved from di n p on the

841
01:14:10,409 --> 01:14:17,130
lc-ms I'm sorry

842
01:14:17,460 --> 01:14:28,170
yeah it's reverse phased yep and and
this slide is just kind of giving you a

843
01:14:28,170 --> 01:14:34,950
side-by-side a di NP in the upper right
on the lc/ms di NP and the lower left

844
01:14:34,950 --> 01:14:43,290
and the GCMs that you can see it's hard
to see of course because di NP n GCMs

845
01:14:43,290 --> 01:14:50,160
relative to the single analytes it just
kind of it's not very responsive because

846
01:14:50,160 --> 01:14:56,940
it's just a nice American mix so that's
one advantage that we found with using

847
01:14:56,940 --> 01:15:08,310
lc-ms is really for for those finger
peaks so our approach like I said we use

848
01:15:08,310 --> 01:15:15,300
both GCMs and lc-ms in conjunction with
one another if the lab decides to use

849
01:15:15,300 --> 01:15:21,510
GCMs primarily it's kind of a flow chart
showing what should be done if you find

850
01:15:21,510 --> 01:15:29,400
the family's DN o pd ID PDI NP they
should be confirmed on lc-ms just just

851
01:15:29,400 --> 01:15:36,300
to ensure a good quantitation and just
to confirm that's out there now if

852
01:15:36,300 --> 01:15:43,980
you're going to primarily use LCMS you
can do that as well but then yet there's

853
01:15:43,980 --> 01:15:48,360
things you need to know about that
though if you find DB p dibutyl

854
01:15:48,360 --> 01:15:54,090
phthalate which is regulated it also
that well it will call a loop with an

855
01:15:54,090 --> 01:15:59,480
unregulated thell a die isobutyl thely
so if you're getting a detection on DB p

856
01:15:59,480 --> 01:16:05,940
it may really be die isobutyl faily
they they essentially Khoa loot with one

857
01:16:05,940 --> 01:16:13,440
another so that's why if you do find DB
p it's it's pretty recommended pretty

858
01:16:13,440 --> 01:16:17,790
urgent that you confirm it by GCMs to
see is it the regulated form of the

859
01:16:17,790 --> 01:16:26,160
unregulated or a mixture of both and
we've seen all three cases there so that

860
01:16:26,160 --> 01:16:32,110
that's pretty much the flow that
that we found to work for us

861
01:16:33,560 --> 01:16:39,350
so in conclusion a dual instrument
system works for us it's been a useful

862
01:16:39,350 --> 01:16:44,960
approach enhancing our data integrity
with challenging samples and like I said

863
01:16:44,960 --> 01:16:51,230
before as time goes on we just just when
you think you've seen it all you haven't

864
01:16:51,230 --> 01:16:56,510
we've seen new plastic plastics come out
with different compounds and adding new

865
01:16:56,510 --> 01:17:04,040
challenges it's like an ongoing process
the learning never ends so this is what

866
01:17:04,040 --> 01:17:07,580
Bureau of our Bureau Bureau Veritas has
found to work for us in our long history

867
01:17:07,580 --> 01:17:12,770
of fail a testing in order to provide
clear results from an otherwise messy

868
01:17:12,770 --> 01:17:18,050
scan just because of this we feel much
more confident and the data that we

869
01:17:18,050 --> 01:17:23,800
issue basically we were having two
detectors give us data if need be

870
01:17:23,800 --> 01:17:30,110
it's also important to note that you
don't need a state-of-the-art gc-ms or

871
01:17:30,110 --> 01:17:35,810
lc/ms I mean you can use models that
been around for years it's not like you

872
01:17:35,810 --> 01:17:39,740
have to go out and buy the the most
sensitive instrument to take this

873
01:17:39,740 --> 01:17:44,870
approach on so that in that sense it
helps labs that I'm looking for you know

874
01:17:44,870 --> 01:17:51,320
cost savings of course so in the end
it's all about just feeling confident

875
01:17:51,320 --> 01:17:58,160
about the results and data that your lab
is issuing with these tricky samples are

876
01:17:58,160 --> 01:18:03,910
there any questions yes

877
01:18:10,400 --> 01:18:17,610
yes the question was do our labs in Asia
used the dual method approach I'm not

878
01:18:17,610 --> 01:18:21,420
sure I from my understanding I know they
they have both okay

879
01:18:21,420 --> 01:18:26,670
my consultant they're saying yes they do
I knew they had both instrumentations

880
01:18:26,670 --> 01:18:31,010
but I just got confirmation that yes
they do

881
01:18:39,969 --> 01:18:46,360
the question was is this our standard
practice in testing for di MP & di DP

882
01:18:46,360 --> 01:18:50,499
yes it is
yeah just because it's just so tricky on

883
01:18:50,499 --> 01:18:55,239
GCMs we wouldn't just we just wouldn't
feel that confident issuing results if

884
01:18:55,239 --> 01:18:58,989
we get something that would really
challenge the chromatography like we

885
01:18:58,989 --> 01:19:06,939
often see on GCMs so yeah this this is
our our good practice here bébé yep go

886
01:19:06,939 --> 01:19:09,059
ahead

887
01:19:16,480 --> 01:19:25,670
on LCMS sure actually I can go back
that's fine um actually if we overlaid

888
01:19:25,670 --> 01:19:30,830
them it show better but this is there
are two separate identic stretches as

889
01:19:30,830 --> 01:19:36,350
well this is a leaf for tickets for 19
I'm the LCMS you want to take a look at

890
01:19:36,350 --> 01:19:42,080
this you know starting off it you know
ten and a half going to twelve and here

891
01:19:42,080 --> 01:19:52,430
we got twelve - like 14 for di DPL CMS
total run time including the post run

892
01:19:52,430 --> 01:19:58,900
and leave just 22 minutes something like
that 24 minutes

893
01:20:08,410 --> 01:20:25,700
yeah yep okay yeah we're involved in
internal correlations with their own

894
01:20:25,700 --> 01:20:29,600
company as well as external correlations
to see if the methodologies that other

895
01:20:29,600 --> 01:20:33,410
labs are using are comparable with ours
so yeah that's that's part of our

896
01:20:33,410 --> 01:20:41,080
quality control
all right let's thank Chris thank you

897
01:20:42,380 --> 01:20:46,780
up next we have Luke Ackerman from the
FDA

898
01:20:57,180 --> 01:21:05,290
thank you all for coming and thanks for
organizing this okay okay so today I'm

899
01:21:05,290 --> 01:21:10,120
going to be discussing dart MS or direct
analysis in real time which is an

900
01:21:10,120 --> 01:21:15,340
ambient ionization mass spectrometry
technique as a possible screening method

901
01:21:15,340 --> 01:21:18,910
for phthalates specifically we've been
doing a lot of work on food-contact

902
01:21:18,910 --> 01:21:25,239
polymers and looking for rapid screening
techniques phthalates just happens to be

903
01:21:25,239 --> 01:21:31,260
one of the model additives that we look
at whenever we evaluate methodology so I

904
01:21:31,260 --> 01:21:36,460
thought perhaps I could share a little
bit of our work with you here and I

905
01:21:36,460 --> 01:21:40,510
think just to kind of give you the
conclusion I think dart can be used as a

906
01:21:40,510 --> 01:21:46,060
very effective screening technique it
has limited capabilities otherwise but

907
01:21:46,060 --> 01:21:50,410
I'll be going over the details of that
and I work at the Center for Food Safety

908
01:21:50,410 --> 01:21:56,469
and applied nutrition at the FDA we're a
research arm of the FDA so I'm not

909
01:21:56,469 --> 01:22:02,110
conducting any regulatory enforcement
actions at our facility we're developing

910
01:22:02,110 --> 01:22:06,760
methodology and trying to advance the
science at the agency so I won't be

911
01:22:06,760 --> 01:22:14,140
discussing any legal matters either so
how does dart work well dart is as I

912
01:22:14,140 --> 01:22:19,180
said an ambient ionization technique
it's just one commercial application of

913
01:22:19,180 --> 01:22:26,100
many different you've probably heard of
desi dart a sap gosh what else nano desi

914
01:22:26,100 --> 01:22:32,110
supersonic electrospray ionisation these
are all the academic terms for various

915
01:22:32,110 --> 01:22:38,050
techniques of producing gas phase ions
off of condensed materials so solids and

916
01:22:38,050 --> 01:22:43,540
liquids and turning them into gas phase
ions such that mass spectrometry can be

917
01:22:43,540 --> 01:22:48,910
used to help analyze the stuff in
question in this particular case dart is

918
01:22:48,910 --> 01:22:52,930
unique and was probably granted a patent
because it quenches a plasma it

919
01:22:52,930 --> 01:22:57,400
generates a plasma of helium in the glow
discharge area of the diagram here and

920
01:22:57,400 --> 01:23:02,560
then it quenches them on the grounded
electrodes in a further down stream it

921
01:23:02,560 --> 01:23:07,750
then subsequently heat it heats that gas
and the quenched plasma generates

922
01:23:07,750 --> 01:23:12,820
excited helium or metastable helium
which is just an electron shifted up in

923
01:23:12,820 --> 01:23:18,880
energy level in the orbitals of helium
once you heat this helium stream you now

924
01:23:18,880 --> 01:23:23,649
have hot excited helium it exits the
dart source and into the atmosphere of

925
01:23:23,649 --> 01:23:28,809
the lab where you place your sample the
atmosphere in the lab not surprisingly

926
01:23:28,809 --> 01:23:35,499
contains water vapor and water has a
good match with the energy of the

927
01:23:35,499 --> 01:23:40,510
excited helium and it produces
protonated water clusters and these

928
01:23:40,510 --> 01:23:45,489
protonated water clusters in turn
ionized any gas phase molecules in the

929
01:23:45,489 --> 01:23:51,219
vicinity of the source there's been lots
of studies on how big of a gap you need

930
01:23:51,219 --> 01:23:56,019
and for different compounds different
distances between the exit of the dart

931
01:23:56,019 --> 01:23:58,899
and the entrance of the mass
spectrometer will yield better results

932
01:23:58,899 --> 01:24:04,119
but in general your protonating whatever
gas phase ions you can get and the

933
01:24:04,119 --> 01:24:10,689
thermal heating of the helium gas also
allows thermal transfer to the surfaces

934
01:24:10,689 --> 01:24:15,369
so you're doing thermal desorption so
dart is thermal desorption atmospheric

935
01:24:15,369 --> 01:24:19,689
chemical ionization similar to the
negative and positive chemical

936
01:24:19,689 --> 01:24:24,669
ionization possibilities with standard
GCMs techniques but this is done in the

937
01:24:24,669 --> 01:24:28,899
open lab atmosphere and the advantage of
that is that allows you to stick odd

938
01:24:28,899 --> 01:24:32,649
shaped objects in front of a mass
spectrometer so you don't have to do

939
01:24:32,649 --> 01:24:37,570
sample preparation in order to get some
sort of representative spectra of the

940
01:24:37,570 --> 01:24:42,909
surfaces of the samples that you're
looking at as you can imagine that

941
01:24:42,909 --> 01:24:47,380
creates a whole host of other questions
as what kinds of surfaces how close what

942
01:24:47,380 --> 01:24:51,280
compounds how hot those are all
parameters that have to be dictated

943
01:24:51,280 --> 01:24:55,899
according to each application but this
is the basic approach that dart takes to

944
01:24:55,899 --> 01:25:02,199
producing gas phase ions for mass
spectrometry and what does Dart actually

945
01:25:02,199 --> 01:25:06,820
look like in practice we've generated a
little movie here one of the advantages

946
01:25:06,820 --> 01:25:12,929
of dart is how fast it is and this will
show you a mass spectrum of a little

947
01:25:12,929 --> 01:25:17,169
sample that was introduced you probably
maybe saw in an upper right hand corner

948
01:25:17,169 --> 01:25:22,090
a little gap glass capillary came down
and was stuck into the dart stream and

949
01:25:22,090 --> 01:25:26,440
you see a peak there at M over Z one
seven this was our demonstration of how

950
01:25:26,440 --> 01:25:32,260
melamine is detected in milk products so
we coated a little glass capillary with

951
01:25:32,260 --> 01:25:36,400
a milk product and if there was melamine
present it was Heian eyes dan M over Z

952
01:25:36,400 --> 01:25:42,639
127 mass of melamine plus h or plus
proton jumps up on the screen there and

953
01:25:42,639 --> 01:25:47,440
you can see the time frame there is on
the order of 30 seconds or less per

954
01:25:47,440 --> 01:25:54,090
sample so you get an immediate spike in
signal and it drops off rather quickly

955
01:25:54,090 --> 01:25:59,230
as fast as your robot can go and pick up
another sample or you can go grab

956
01:25:59,230 --> 01:26:03,239
another sample it returns to baseline
and you're ready for the next analysis

957
01:26:03,239 --> 01:26:09,280
so what does dart m/s measure well it
measures mass spectrum because it's dart

958
01:26:09,280 --> 01:26:13,900
is just an ionization TKE technique for
a mass spectrometer in this case we

959
01:26:13,900 --> 01:26:20,500
interface dart with a Watters última
which is just a simple triple quad and

960
01:26:20,500 --> 01:26:25,510
we looked at the mass spectrum because
of the way that we set up the dart and

961
01:26:25,510 --> 01:26:29,860
because of the mass spec you can see a
couple of diagnostic ions here for in

962
01:26:29,860 --> 01:26:35,349
this case diethyl hexyl phthalates you
see the molecular ion the n plus h some

963
01:26:35,349 --> 01:26:38,199
of the characteristic fragment ions we
were running this under some slightly

964
01:26:38,199 --> 01:26:43,360
fragmenting conditions and if you use
different mass spectrometers you get

965
01:26:43,360 --> 01:26:46,929
different results so the bottom trace
here is the mass spectrum of the same

966
01:26:46,929 --> 01:26:52,179
standard using a Joel acute off which is
just a low-end time-of-flight mass

967
01:26:52,179 --> 01:26:58,510
spectrometer about 170 K for one of
those and as you can see there even in

968
01:26:58,510 --> 01:27:03,429
the presence of a mix standard this
happened to have a little di n pol di DP

969
01:27:03,429 --> 01:27:09,460
and some teh P you can see the dominant
ion because of the chyme on the trace

970
01:27:09,460 --> 01:27:16,570
that I picked was that well the octal
phthalate I should say signal and the

971
01:27:16,570 --> 01:27:21,400
nice thing about using a toss is of
course the added mass accuracy as you

972
01:27:21,400 --> 01:27:27,130
can see you get high mass accuracy M to
a certain degree a good mass resolution

973
01:27:27,130 --> 01:27:30,070
there's certainly other mass
spectrometers that can achieve better

974
01:27:30,070 --> 01:27:35,700
but the point being that when you're
doing direct analysis or atmospheric is

975
01:27:35,700 --> 01:27:40,560
analysis you're gonna get mixtures and
you no longer are using chromatography

976
01:27:40,560 --> 01:27:45,150
to help separate your chemicals previous
to analysis so anything that you can do

977
01:27:45,150 --> 01:27:49,380
to help you differentiate one molecule
from another can help and accurate and

978
01:27:49,380 --> 01:27:54,390
high res mass spec can help with that
so what does the dart ms response look

979
01:27:54,390 --> 01:27:57,060
like well you're not using
chromatography so you don't have a

980
01:27:57,060 --> 01:28:02,370
chromatogram now you just have a time
trace this is a typical total ion

981
01:28:02,370 --> 01:28:05,760
current it's not a total item
chromatogram there's no chromatography

982
01:28:05,760 --> 01:28:13,080
of a couple of sample introductions in
front of a dart and in this case this is

983
01:28:13,080 --> 01:28:18,630
from a triple quad the waters that I was
mentioning and you see the TIC rise and

984
01:28:18,630 --> 01:28:23,820
fall but rather noisily and rather air
reproducibly but when you go to look at

985
01:28:23,820 --> 01:28:28,820
a particular compound of interest you
see that the trace is a lot more easy to

986
01:28:28,820 --> 01:28:35,640
understand and in this case the the dhp
standard of various concentrations gave

987
01:28:35,640 --> 01:28:40,200
a very predictable response as a
function of time so again we're looking

988
01:28:40,200 --> 01:28:45,090
at a mass spectrum as a function of time
so what are all the different dark

989
01:28:45,090 --> 01:28:50,010
configurations that I've been discussing
well we started off in our laboratory

990
01:28:50,010 --> 01:28:54,180
with a first generation dart which is
kind of a clunky looking thing and in

991
01:28:54,180 --> 01:28:58,560
this particular case we are using a
robotic auto sampler to place little

992
01:28:58,560 --> 01:29:02,640
strips of particular polymers into the
dart stream you'll see a little glass

993
01:29:02,640 --> 01:29:07,890
capillary tube sticking out that helps
transfer the gas stream from the

994
01:29:07,890 --> 01:29:10,890
ionization region into the mass
spectrometer and because the

995
01:29:10,890 --> 01:29:16,500
manufacturer of the dart ion source had
not interfaced this particular mass

996
01:29:16,500 --> 01:29:20,190
spectrometer we literally just
terminated the glass tube immediately

997
01:29:20,190 --> 01:29:25,230
adjacent to the mass spectrometer Inlet
it wasn't even on axis and so in that

998
01:29:25,230 --> 01:29:29,820
case our sensitivity was quite poor and
the results that we generate with this

999
01:29:29,820 --> 01:29:33,120
configuration weren't really
translatable to a lot of the people who

1000
01:29:33,120 --> 01:29:38,370
had purchased the Jol acute off switch
was the parent company of the ion source

1001
01:29:38,370 --> 01:29:44,360
manufacturer dart is manufactured by
iron sights

1002
01:29:44,360 --> 01:29:50,060
out of p body mass we subsequently
purchased two other configurations at

1003
01:29:50,060 --> 01:29:55,730
dart Orbitrap in a dart a cute off so
this is the joel a cute off and the

1004
01:29:55,730 --> 01:30:00,260
newest version of their dart source that
goes with the a cute off again as you

1005
01:30:00,260 --> 01:30:05,750
can see here the the silver cone coming
out of the mass spectrometer that's the

1006
01:30:05,750 --> 01:30:10,340
inlet the silver and white cones coming
off the dart source the that's where the

1007
01:30:10,340 --> 01:30:15,650
gas exits we happen to be running a
little homemade rail linear rail through

1008
01:30:15,650 --> 01:30:19,040
the dart source at this particular case
we were mapping some chemical

1009
01:30:19,040 --> 01:30:23,960
concentrations across the surface of a
piece of packaging and we just thought

1010
01:30:23,960 --> 01:30:27,500
we'd try a little linear rail in that
particular case but as you can see the

1011
01:30:27,500 --> 01:30:30,949
dart source can be moved back and forth
various distances from the mass

1012
01:30:30,949 --> 01:30:34,639
spectrometer Inlet they even have
configurations that allow you to raise

1013
01:30:34,639 --> 01:30:39,590
and lower it angle it at different
angles and so you really can adjust the

1014
01:30:39,590 --> 01:30:43,280
dart source to fit whatever geometries
you need for your particular sand

1015
01:30:43,280 --> 01:30:49,219
samples the beautiful thing about dart
is you can stick really crazy shaped

1016
01:30:49,219 --> 01:30:53,000
objects in front of it and you can still
get reproducible mass spectrum you don't

1017
01:30:53,000 --> 01:30:56,719
have to mill you don't have to
homogenize now you lose all that

1018
01:30:56,719 --> 01:31:02,000
information you lose reproducibility and
quantitation and the like because you're

1019
01:31:02,000 --> 01:31:07,429
not looking at the same sample as you
would with in in a typical regulatory

1020
01:31:07,429 --> 01:31:11,810
analysis but if you're going to screen
stuff and if you want to have an idea of

1021
01:31:11,810 --> 01:31:17,690
what we're on a particular object
phthalates or any other compound might

1022
01:31:17,690 --> 01:31:24,080
be for the surfaces of those objects
then dart will allow you to to do that

1023
01:31:24,080 --> 01:31:28,429
by looking at only one part of the
object at a time or if you're really

1024
01:31:28,429 --> 01:31:32,510
want to cover every surface you could
systematically run the whole surface

1025
01:31:32,510 --> 01:31:38,420
through there but in either case dart is
a quick way for a laboratory worker to

1026
01:31:38,420 --> 01:31:43,280
look at a sample without having to do
prep and without cross-contamination

1027
01:31:43,280 --> 01:31:46,190
because there's nothing to contaminate
here the entrance to the mass

1028
01:31:46,190 --> 01:31:51,770
spectrometer which is going to be less
contaminated than if you inject a plug

1029
01:31:51,770 --> 01:31:58,159
of solvent from from a GC system so it's
it's got some advantages that way

1030
01:31:58,159 --> 01:32:03,530
here's another angle us introducing a
little piece of of packaging you can do

1031
01:32:03,530 --> 01:32:07,790
it manually you can use linear rails
they've got transmission configurations

1032
01:32:07,790 --> 01:32:14,599
where you can deposit extracts onto
little gridded wire meshes and the and

1033
01:32:14,599 --> 01:32:18,860
the helium flows through the wire mesh
and dissolves whatever was on the

1034
01:32:18,860 --> 01:32:23,270
solution that deposited on the grid you
know there's lots of configurations for

1035
01:32:23,270 --> 01:32:28,310
introducing samples via dart so can it
quantitate that's the big question and

1036
01:32:28,310 --> 01:32:33,619
the answer is no so these are some just
some straight-up

1037
01:32:33,619 --> 01:32:39,590
solvent calibration solutions and just
measuring the peak height of the mass

1038
01:32:39,590 --> 01:32:43,070
effect signal for diethyl hexyl file a
very simple task

1039
01:32:43,070 --> 01:32:49,420
not at all linear not very reproducible
at that and the reason is is that

1040
01:32:49,420 --> 01:32:53,780
various amounts of materials are
deposited onto the tip of the glass

1041
01:32:53,780 --> 01:32:58,699
capillary that we use to sample the
liquids in GC you're using a syringe to

1042
01:32:58,699 --> 01:33:04,550
deliver exactly one microliter every
single time whereas in dart the current

1043
01:33:04,550 --> 01:33:08,780
autosampler configuration dips a melting
point capillary into your liquid

1044
01:33:08,780 --> 01:33:14,570
solution and depending on how clean that
capillary is how far it's dipped in how

1045
01:33:14,570 --> 01:33:21,139
quickly you know what was there a drop
shaken off the capillary in the robots

1046
01:33:21,139 --> 01:33:25,010
movements over to the dart you get
various quantities deposited on the

1047
01:33:25,010 --> 01:33:28,790
surface and then how fast you scan it
through there were there any air

1048
01:33:28,790 --> 01:33:33,650
disturbances in the lab that shifted gas
flow patterns it desorbs a different

1049
01:33:33,650 --> 01:33:37,130
amount of material off of that and a
different amount makes it into the mass

1050
01:33:37,130 --> 01:33:41,599
spec and a different ion signal but one
way to account for all those differences

1051
01:33:41,599 --> 01:33:46,010
is internal standards you know in GCMs
we often do that you get very nice

1052
01:33:46,010 --> 01:33:49,579
linear calibration curves when you
internal standard normalize your

1053
01:33:49,579 --> 01:33:55,280
responses can also it also helps you
with troubleshooting issues in your

1054
01:33:55,280 --> 01:34:00,139
methodology recovery correction you
don't have to take you don't have to do

1055
01:34:00,139 --> 01:34:04,969
quantitative transfers all of those
things are helped by internal standards

1056
01:34:04,969 --> 01:34:09,110
and so is dart calibration when you
include just two internal standards in

1057
01:34:09,110 --> 01:34:12,949
this case for these suite of
Leites that you see on the bottom right

1058
01:34:12,949 --> 01:34:17,900
I included a labeled standard of diethyl
phthalate and a label of diethyl hexyl

1059
01:34:17,900 --> 01:34:23,749
phthalates and then normalized to one of
those to the relative response is quite

1060
01:34:23,749 --> 01:34:30,709
linear and you know R Squared's that we
go on about all better than 0.995 but

1061
01:34:30,709 --> 01:34:35,479
when you look at them on a log scale you
see GCMs is still much more linear than

1062
01:34:35,479 --> 01:34:40,880
dark and that makes sense
your just a lot more controlled sample

1063
01:34:40,880 --> 01:34:48,650
introduction a lot more controlled ion
transfer into the inlet not to say that

1064
01:34:48,650 --> 01:34:52,459
it can't be useful you can't get good
semi quantitative specially out of

1065
01:34:52,459 --> 01:34:57,050
liquid solutions but it's never going to
be the same as GCMs and i would like to

1066
01:34:57,050 --> 01:35:00,949
point out this is over five orders of
magnitude so clearly a very broad range

1067
01:35:00,949 --> 01:35:08,689
here and and but it's still GCMs is the
gold standard for quantitation so it can

1068
01:35:08,689 --> 01:35:13,429
quantitate in solutions but we're not
looking at solutions we're looking at

1069
01:35:13,429 --> 01:35:18,590
surfaces so how sensitive is dart when
it comes to looking at surfaces well we

1070
01:35:18,590 --> 01:35:25,249
had another project where we were
looking at photo initiators transferring

1071
01:35:25,249 --> 01:35:29,539
from the print side of a printed piece
of food packaging to the food contact

1072
01:35:29,539 --> 01:35:34,610
side of an adjacent piece of food
packaging it's called set-off and we

1073
01:35:34,610 --> 01:35:38,869
decided to look at err geekier 180 for a
common photo initiator by producing

1074
01:35:38,869 --> 01:35:43,880
standard samples with various levels of
the photo initiator on the surface of

1075
01:35:43,880 --> 01:35:50,840
the polymer this I believe was in LDPE
film and what we saw was just some of

1076
01:35:50,840 --> 01:35:55,329
the concentrations of the samples that
we prepared you can see that the

1077
01:35:55,329 --> 01:36:00,260
surficial concentrations at dart which
dart is able to differentiate from the

1078
01:36:00,260 --> 01:36:06,739
blanks is right around 0.25 nanograms
per square centimeter so definitely down

1079
01:36:06,739 --> 01:36:14,989
in the low fractions of a hundredths of
a percent level you're going to see

1080
01:36:14,989 --> 01:36:18,979
surficial concentrations orders of
magnitude higher if you're looking at

1081
01:36:18,979 --> 01:36:25,190
even 0.1% phthalates in a polymer so
clearly it's

1082
01:36:25,190 --> 01:36:28,520
more than sensitive enough for these
types of compounds we haven't done this

1083
01:36:28,520 --> 01:36:33,650
exact procedure with phthalates but the
responsiveness of dart and area 2 of

1084
01:36:33,650 --> 01:36:37,400
dart to area cure and phthalates is very
comparable they're both small molecules

1085
01:36:37,400 --> 01:36:40,520
I think this vapor pressure of this
compound is almost exactly the same as

1086
01:36:40,520 --> 01:36:44,660
diethyl hexyl phthalates vapor pressure
is a pretty big factor when you're

1087
01:36:44,660 --> 01:36:49,130
thermally desorbing something off of a
surface so we see pretty comparable

1088
01:36:49,130 --> 01:36:53,690
responses in that case i expensed expect
the sensitivity for phthalates to be

1089
01:36:53,690 --> 01:36:58,699
very comparable and then we also need
the technique to identify the phthalates

1090
01:36:58,699 --> 01:37:03,199
and in particular the nice thing about
dart as I showed you back at the

1091
01:37:03,199 --> 01:37:07,070
beginning is it produces almost
exclusively molecular ion that is M plus

1092
01:37:07,070 --> 01:37:10,730
h you know one of the reasons why we
like to use chemical ionization it's a

1093
01:37:10,730 --> 01:37:14,810
much softer ionization technique it
doesn't give you as reproducible mass

1094
01:37:14,810 --> 01:37:20,930
spectrum on those fragment ions but it
gives you a lot of very specific

1095
01:37:20,930 --> 01:37:26,690
molecular ion to work with you can also
include a little ammonia hydroxide in in

1096
01:37:26,690 --> 01:37:30,440
the lab environment or swab it onto the
surface of a polymer and you can produce

1097
01:37:30,440 --> 01:37:37,489
ammonia instead of protonated molecule
ions so instead of m plus h SN + NH 4

1098
01:37:37,489 --> 01:37:41,510
it's often a common technique to look at
both of those ions as a way of

1099
01:37:41,510 --> 01:37:48,590
confirming that you know the 195 iron
you see is really corresponding to to

1100
01:37:48,590 --> 01:37:54,860
dimethyl phthalates or whatnot and as we
all know this m plus h ion is very

1101
01:37:54,860 --> 01:37:59,690
useful the fragment ions help a little
bit more but isomers produce all the

1102
01:37:59,690 --> 01:38:03,469
same ions in theory and they should
because there i Cerner's they're

1103
01:38:03,469 --> 01:38:07,880
positional isomers so diethyl hexyl
phthalates and octal phthalates are

1104
01:38:07,880 --> 01:38:14,719
perfect example and other work by other
folks including Roth and Bakker and and

1105
01:38:14,719 --> 01:38:19,489
SH walk it published in rapid
communication and mass spec took a close

1106
01:38:19,489 --> 01:38:23,600
look at this I think cooky out of
Hungary also did the same thing in

1107
01:38:23,600 --> 01:38:31,340
international mass spec Journal and what
they found is the 261 to 279 the the

1108
01:38:31,340 --> 01:38:37,280
Ethne minus ethylene and minus ethanol
fragments are produced in differential

1109
01:38:37,280 --> 01:38:40,780
ratios
and that really froze up and you can

1110
01:38:40,780 --> 01:38:45,250
differentiate ethyl hexlen and octal
that way the trouble is a lot of things

1111
01:38:45,250 --> 01:38:52,600
can produce 261 and 279 and pretty much
any isomeric octal phthalates can do

1112
01:38:52,600 --> 01:38:56,949
that to varying degrees and what if you
have mixtures of these phthalates so

1113
01:38:56,949 --> 01:39:02,500
when we looked at a particular PP and
LDPE film for various phthalates we see

1114
01:39:02,500 --> 01:39:07,449
the molecular ions of ethyl Hextall or
an octal and also dye I so know no

1115
01:39:07,449 --> 01:39:12,910
phthalates when we look at the iron
ratios of the fragments 279 clearly it

1116
01:39:12,910 --> 01:39:19,750
exceeds 261 suggesting it's ethyl hexo
phthalates you know the ratio is 1.7 but

1117
01:39:19,750 --> 01:39:25,270
in the standards that ratio was 3.3 so
what does this mean and in our initial

1118
01:39:25,270 --> 01:39:29,170
analysis of this sample which is the
cookie bar at the bottom of the chart

1119
01:39:29,170 --> 01:39:34,000
there we said that we thought there's
low hundreds of nanograms per square

1120
01:39:34,000 --> 01:39:38,680
centimeter of diethyl hexyl phthalates
and we didn't think there was an octal

1121
01:39:38,680 --> 01:39:44,050
phthalates we then subjected these same
samples to extraction LCMS analysis

1122
01:39:44,050 --> 01:39:49,000
here's the chromatogram and we focused a
lot on that isononyl phthalates in that

1123
01:39:49,000 --> 01:39:55,690
particular case and what we saw was a
very odd shape lcms peak we expect nice

1124
01:39:55,690 --> 01:40:00,820
sharp lcms Peaks for the ice and onal
and we suspected it was an N octave

1125
01:40:00,820 --> 01:40:06,460
Dessel phthalates which would be a
particular structural isomer of of the

1126
01:40:06,460 --> 01:40:10,780
no no phthalates and sure enough when we
spiked it with an additional hundred ppb

1127
01:40:10,780 --> 01:40:16,719
of dinnah that small shoulder goes up so
din up is the shoulder and n octave

1128
01:40:16,719 --> 01:40:20,320
Dessel phthalates was the dominant
compound in this particular package

1129
01:40:20,320 --> 01:40:26,020
similar analysis of ethyl hexyl and an
octal in this particular and compound

1130
01:40:26,020 --> 01:40:30,250
the cookie bar showed as you can see on
the bottom that there was actually

1131
01:40:30,250 --> 01:40:34,870
nearly equivalent concentrations of an
octal phthalates and ethyl hexyl

1132
01:40:34,870 --> 01:40:39,460
phthalates that's why our iron ratios of
those fragments switched by dart ms and

1133
01:40:39,460 --> 01:40:45,180
we completely missed the anak the daesil
phthalates in this particular sample so

1134
01:40:45,180 --> 01:40:50,140
what can we can conclude from all of
this well sample positioning effects

1135
01:40:50,140 --> 01:40:54,550
dart response it's
more than sensitive enough for phthalate

1136
01:40:54,550 --> 01:40:59,559
analysis it can be quantitative in
certain situations when you use internal

1137
01:40:59,559 --> 01:41:03,309
standards and especially with solvents
it can be sending semi-quantitative off

1138
01:41:03,309 --> 01:41:07,510
of polymer surfaces especially if you
can deposit an internal standard which I

1139
01:41:07,510 --> 01:41:13,329
know some nist groups have been working
on for the explosives methods they've

1140
01:41:13,329 --> 01:41:18,849
been working on printing standards on
surfaces and it cannot reliably identify

1141
01:41:18,849 --> 01:41:25,420
different isomers of different
phthalates not in mixtures and one last

1142
01:41:25,420 --> 01:41:31,690
note is the time responsive dart so this
is a typical tick and this is what that

1143
01:41:31,690 --> 01:41:38,440
stand this is a total iron current of a
typical standard for Dart and what you

1144
01:41:38,440 --> 01:41:43,570
see here is first the solvent ethyl
hexyl phthalates or ethyl acetate excuse

1145
01:41:43,570 --> 01:41:49,719
me came off the surface of the of the
tip and then the diethyl hexyl

1146
01:41:49,719 --> 01:41:54,130
phthalates in the last second and the
time scales of these are about 1/6 to

1147
01:41:54,130 --> 01:41:59,199
1/9 of a second so you need to be
measuring your your mass spectral traces

1148
01:41:59,199 --> 01:42:05,559
at 6 to 9 Hertz at least so cycles per
second in order for dart to really be

1149
01:42:05,559 --> 01:42:10,570
useful in capturing the actual quantity
law so these are all things that we need

1150
01:42:10,570 --> 01:42:14,289
to keep in mind but again the main
points it cannot reliably identify the

1151
01:42:14,289 --> 01:42:17,920
different isomers it can quant under
certain circumstances but it can

1152
01:42:17,920 --> 01:42:24,269
certainly screen for the presence of
various phthalates thank you

1153
01:42:32,160 --> 01:42:40,060
well if you can reliably move the sample
through the dart stream like we were

1154
01:42:40,060 --> 01:42:45,280
doing with that little linear rail on
the surface then you can assign a

1155
01:42:45,280 --> 01:42:50,140
spatial map that will be our next
publication where we reproduce the set

1156
01:42:50,140 --> 01:42:54,960
off image of a print on the adjacent
piece of packaging you can see the same

1157
01:42:54,960 --> 01:43:09,090
trademark logos and you know images but
just recreate it in a chemical map well

1158
01:43:09,090 --> 01:43:16,200
so the width of the of the dart beam is
on the order of about a centimeter or

1159
01:43:16,200 --> 01:43:20,500
three-quarters of a centimeter somewhere
around there I know that they're working

1160
01:43:20,500 --> 01:43:26,880
on different iterations that have
narrower helium being widths but really

1161
01:43:26,880 --> 01:43:30,700
double or triple that and that's
probably the double that and that's

1162
01:43:30,700 --> 01:43:33,610
probably the smallest scale of
resolution you're going to achieve and

1163
01:43:33,610 --> 01:43:37,530
then of course you can screw it up by
moving the sample too fast or you're

1164
01:43:37,530 --> 01:43:54,790
reproducibly yeah why don't we go ahead
we'll break for lunch now and we'll

1165
01:43:54,790 --> 01:43:58,290
start back up again at one o'clock

